7.1

SPECIFICATIONS

Reference

: "Hydro-Wiper A" from ZEBRA-TECH

Wiping interval

:

15, 30, 45, 60, 120, 180, 240, 300, 360 or 720 minutes

Wiping time lag:

+/- 1 minute per year (0 – 40°C)

Power supply:

6 AA alkaline batteries (DURACELL MN1500 battery or equivalent)

Depth limit:

30 m (optional 100 m)

7.1.1

Assembly

Remove the protective cap from turbidity sensor

-

Unscrew the two screws HC

-

Remove the protective cover

Insert the STBD probe in the flange

-

Remove the screw of the STBD probe

-

Insert the Turbidity sensor first

-

Focus  the  probe  so  that  the  temperature  sensor  passes
in the open window

-

Position  the  sensor  so  that  the  brush  covers  the  optical  window
for moving.

-

Secure the sensor by tightening the fixing screw.

-

Adjust the pressure of the brush to playing on the length of the
arm

For more information, please refer to the user guide from ZEBRA-TECH

Sens de montage

Содержание STBD-PR

Страница 1: ...w nke instrumentation com Tel 02 97 36 41 31 Fax 02 97 36 55 17 PAGE 1 18 USER MANUAL STBD PR Rev 5 Document reference STBD 33 30 065_UTI_UK doc Document applicable to product s STBD PR station and pr...

Страница 2: ...JUSTMENT TEMPERATURE 10 5 4 TURBIDITY SENSOR 11 5 4 1 Adjustment 11 6 TECHNICAL CHACTERISTICS 13 6 1 MECHANICAL CHARACTERISTICS 13 6 2 SPECIFICATIONS OF RADIO TRANSMISSION 13 6 3 MEMORY AUTONOMY 13 6...

Страница 3: ...ts its data to a CONCENTRATOR via a radio link Applications Suspended biomass visibility pollution sediment transport particle profile Equipment required for the use of an STBD PR 1 Data Pencil readin...

Страница 4: ...http www nke instrumentation com Tel 02 97 36 41 31 Fax 02 97 36 55 17 PAGE 4 18 2 PHOTO Turbidity sensor Seapoint Temperature sensor Pressure sensor Protective cover Shoulders for attaching...

Страница 5: ...automatic mode the probe selects the best range to maximize the accuracy of measurement The algorithm that allows automatic range selection carries out 1 to 3 measurements to determine the optimum ran...

Страница 6: ...Maximum resolution 11m C at 0 C 13m C at 10 C 20m C at 20 C Accuracy 50m C in the range 0 C 20 C Reponse time STBD PR 0 5s at 63 Reponse time STBD PR 2s at 63 3 2 4 Pressure or depth channel Pressure...

Страница 7: ...ies the radio address of the data logger unique factory defined address for each product The Radio Data Pencil uses this address to communicate with the STBD PR data logger Address of the Concentrator...

Страница 8: ...SENTATION OF THE CONCENTRATOR Designed to be installed on board a boat it receives the measurement files transmitted by data loggers such as the STBD PR via the radio connection It also enables the lo...

Страница 9: ...ke a Play Panel and select Channel 0 in the channels configuration tab Raise the calibration coordinates example Vcan0 point s of the converter correspond to the physical value P0 Vcan1 point s of the...

Страница 10: ...t throughout the recording thermostatic bath recommended Read the STBD once again and save the measurements to a file Display the temperature curve and record the raw value measured for each bath This...

Страница 11: ...750 NTU 4 Automatic 5 Select the maximum resolution i e 12 bits for the turbidity channel Channel 2 Prepare two baths of upper and lower calibration for each range Example Preparation of 2 baths of 5...

Страница 12: ...nt s of the converter correspond to the physical value TX Y point s of the converter correspond to the physical value TY Display the file to check the turbidity values of the different baths Edit the...

Страница 13: ...OMY The memory autonomy indicated thereafter is provided for the operating mode Manual start and stop with an interval between each measurement estimated at 8 points of the resolution 6 3 1 STBD PR in...

Страница 14: ...nction of the following parameters in order of significance 1 Measurement frequency or average of the various frequencies used 2 Start stop mode used e g A mode that uses a start or stop on condition...

Страница 15: ...te by the Winmemo II software To help you determine the energy capacity and remaining life time the Winmemo II software provides an estimate of this information based on the average working mode of th...

Страница 16: ...be s field of view is likely to be damaged If the brush is pressed with insufficient force it will not clean effectively Brush pressure can be adjusted by removing the nut at the end of the wiper arm...

Страница 17: ...mbly Remove the protective cap from turbidity sensor Unscrew the two screws HC Remove the protective cover Insert the STBD probe in the flange Remove the screw of the STBD probe Insert the Turbidity s...

Страница 18: ...FOR RADIO OPERATION The radio antenna is positioned on the side of the CTN bracket and fields of view of the Turbidity sensor For maximum radio range the STBD PR must be free of any mechanical obstru...