Instruction manual
NivuSonic Clamp on
®
Fig. 8-73
Selecting the analog output simulation
Channel number
Choose the analog output which is to be simulated by entering the respective
number from 1 – 4 or by using the >left< or >right< arrow keys in the simulation
main menu.
Simulation
Enter the desired value in mA and confirm with Enter in order to directly output it
on the according clamp.
Fig. 8-74
Executing the simulation
Digital outputs
Select the desired relays to be simulated by pressing the >left< or >right< arrow
keys. The relay number chosen is going to be indicated in the first line of the
output table.
The arrow keys >up< or >down< will either directly enable or disable the relay
selected previously.
After selecting >Digital Outputs< the PIN must be entered once again. This en-
sures to keep unauthorised persons away from executing simulations during
operation.
Fig. 8-75
Relay simulation
NivuSonic CO - Rev. 00 as of 08.12.2010
Page 83