Shut-off valves and control fittings shall be installed
always downstream
of flow
velocity sensors.
Fig. 6-12
Using shut-off valves and control fittings
6.3.3 Sensor Connection
The sensors are
available with
permanently attached cable. It is not allowed to
extend or shorten the sensor cable. Connect individual path sensors directly to
the NivuChannel transmitter (one or two sensor pairs). Use only the dedicated
cable provided by NIVUS for connection and / or extension purposes.
The sensor cable can be ordered from NIVUS using the respective article no.
(see ).
It is not allowed to use common extensions in case of different applications or
to use a common signal cable to extend separate level and flow velocity
measurements.
Connect sensor cables to the transmitter in the sensor terminal clamp strip. The
following wiring diagram applies when connecting clamp on sensors:
Fig. 6-13
Connection of sensor 1 to the NivuSonic CO