Chapter 3 Operation
3.12 Pillar Illuminator 100W Operation
78
3.12
Pillar Illuminator 100W Operation
This section describes operations specific to TI-DH Dia Pillar Illuminator 100W.
3.12.1
Condenser Refocusing Clamp
Figure 3-38 Pillar Illuminator 100W
Form the field diaphragm image on the specimen
surface by rotating the condenser focus knob. Rotate
the condenser refocusing clamp clockwise to the
limit to mark this position.
When the condenser is elevated to change the
specimen, it can easily be brought back down to the
initial position (at which the field diaphragm image is
formed) by rotating the condenser focus knob to the
limit. This function is useful for use with high
magnification condensers or when the pillar cannot
be inclined. The condenser refocusing clamp has a
range of motion of 13 mm.
3.12.2
Condenser Mount Rotation
Figure 3-39 Condenser mount
The condenser mount can be rotated if the
condenser mount rotation clamp screw is loosened.
Use this function to adjust the orientation of the turret
when using the DIC attachment.
When using the system condenser without a
polarizer on the condenser holder (i.e. for bright-field
or phase contrast microscopy), this function can also
be used to rotate and fix the turret. This will allow
you to secure a space for manipulator attachment.
Condenser
refocusing clamp
Condenser
focus knob
Condenser
focus knob
Condenser mount
rotation clamp screw
Condenser mount
Содержание Eclipse Ti-E