Chapter 4 Assembly
111
12
Attaching the bottom port (Ti-E/B only)
Figure 4-42 Bottom port adapter attachment
1. Loosen the two bottom port adapter clamp
screws and remove the metal cap from the
bottom port.
2. Insert the bottom port adapter into the
bottom port, and secure it by tightening the
two bottom port adapter clamp screws.
3. Attach adapters to the camera device.
4. Insert the adapters and the camera device
into the bottom port adapter, and secure
them by tightening the two camera adapter
clamp screws.
• Camera devices require compatible adapters
for attachment. First attach the adapters to the
camera device, and then attach the camera
device with the adapters into the bottom port
adapter.
• Attach a protective cap onto the port when the
port is not in use.
Example: Attaching the C mount TV camera to the direct C mount adapter
(1) Screw the C mount TV camera securely into the direct C mount adapter.
(2) Insert the direct C mount adapter into the bottom port adapter, and secure it by tightening the
locking screw.
• When detaching photomicrographic equipment, hold it steady, and then loosen the locking screw.
• If the locking screw is loosened without holding the equipment, the equipment may drop. To avoid
dropping photomicrographic equipment, make sure you have a firm grip before loosening the screw.