Safety Precautions
5
1 Isolate this microscope and its system equipment from the power source during assembly,
connecting/disconnecting cords, lamp replacement, and maintenance
To prevent electric shock and/or malfunctions, always turn off the power switches for the system (press
them to the "O" positions) and unplug the power cords from the wall outlet before assembly, connecting or
disconnecting cords, lamp replacement, and cleaning the microscopy and the objectives.
2 Cautions
when
replacing
lamps
To prevent burn injuries, wait at least 30 minutes after the lamp is turned off to give it sufficient time to cool
down when replacing lamps. And, to avoid electric shock or malfunctions, never attempt to replace lamps
without turning off the power switches for the system (press them to the "O" positions). And then, unplug
the power cords from the wall outlet.
Make sure the lamp house cover is securely fitted to the lamp house after replacing lamps. Never turn on the
lamp while the lamp house cover is open.
When you dispose of the replaced lamp, do not break it up. Instead, dispose of the used lamp as special
industrial waste or dispose of it according to the local regulations and rules.
3 Use the specified light source
Use the specified light source. Using a wrong light source may cause a malfunction or a fire.
•
For the illumination with the optical fiber
Specified light source: Sumita Optical Glass Inc. Halogen light source LS-DWL-N
•
For the illumination with the lamp house
Specified lamp house: Nikon FN-LH
Specified lamp: 12V 100W Long life halogen lamp (PHILIPS 7724i or ODRAM HLX 64623)
Specified power supply: Nikon TE2-PS100W
* In addition, when the lamp house is used for the illumination, a lamp house adapter must be
attached to connect the lamp house with the microscope FN1.
4 Do not spill water or a chemical agent onto the microscope system
Never allow water or a chemical agent to come into contact with the microscope system. And do not use
the microscope system in circumstances where the microscope system is splashed with water or a
chemical agent. Water or a chemical agent splashed onto the power supply or the power supply device for
the optical fiber light source may cause a short circuit, resulting in malfunction or abnormal overheat. If the
product is subject to contact with water or a chemical agent, turn off the power switches for the product
(press them to the "O" positions). Then unplug the power cords from the wall outlet. Then, wipe off the
water with a piece of dry cloth. If water or a chemical agent enters the product, do not use the product and
contact your nearest Nikon representative.
5
Do not place any object on the top of the microscope or its system
Do not place any object on the top of the microscope or its sytem. Do not cover them with a piece of cloth
or so on. The system temperature will rise, resulting in malfunctions.