Chapter 4
Troubleshooting
69
Chapter 4
T
roubleshooting
Problem
Cause
Measure
The entire field of view is
bluish or yellowish.
Filter cube is in the optical path when non
epi-fluorescence observation.
Remove the filter cube from the optical path.
Parts are attached incorrectly.
Confirm that parts (nosepiece, condenser, etc.)
are correctly attached.
(
→
Chapter 3, “2 Assembly for Bright-field
Microscopy”)
Movable parts are not seated correctly.
Correctly set the optical path switching lever,
nosepiece, filter cube switchover turret,
condenser turret, and slider, etc. (Move the
part until it clicks.)
Lack of visibility around
periphery of field of view.
Illumination is uneven
across the field of view.
Field of view is not
visible.
Field diaphragm image is not focused on
the specimen surface.
Make sure the condenser is focused and
centered.
(
→
Chapter 2, “5 Focusing and Centering the
Condenser”.)
Field diaphragm is stopped down too far.
Open the field diaphragm slightly wider than
the field of view.
(
→
Chapter 2 “8 Adjusting the Field Diaphragm”)
Incorrect combination of the objective with
the condenser.
Adopt an appropriate combination.
(
→
Chapter 2 “7 Selecting a Condenser”)
In low magnification microscopy, required
condenser operation was not performed.
Perform the operation such as swinging out or
sliding the condenser.
A lamp is attached incorrectly.
Attach it correctly.
(
→
Chapter 5 “1 Replacing the Lamp (for Ci-S)”)
Lack of visibility around
periphery of field of view.
Illumination is uneven
across the field of view.
Field of view is not
visible.
Lens and specimen are dirty or dusty.
Clean them as appropriate.
(
→
Chapter 5 “2.1 Lense Cleaning”)
The specimen is upside down.
Turn up the cover glass and attach it to the
stage.
(
→
Chapter 2 “2.1 Bright-field Microscopy Procedure
- 5 Place a specimen on the stage, and move the
stage to bring the target into view”)
The thickness of the cover glass is
inadequate.
Attach a cover glass of the specified thickness
(0.17 mm).
Out of focus with an
objective of high
magnification.
Fail-safe device for specimen damage
protection of the objective is pushed in.
Some objective has a stopper to keep the
pushed in state. Turn the tip of the object to
release.
Tip of the objectives without stopper can not
be rotated. Do not use force to draw the
stopper. Contact your nearest Nikon
representative.
An objective is attached incorrectly.
Screw the objective all the way in.
(
→
Chapter 3, “2 Assembly for Bright-field
Microscopy
―
7 Attaching the objective”)
A focal deviation is high
when switching over
objectives.
Diopter adjustment has not been
performed.
Perform diopter adjustment.
(
→
Chapter 2 “4 Adjusting the Diopter”)
The stage is attached incorrectly.
Attach it correctly.
(
→
Chapter 3, “2 Assembly for Bright-field
Microscopy
―
2 Attaching the stage”)
Image is not in focus
although the stage is
raised to the highest
position.
Refocusing position is set lower than the
focusing position.
Check and reset the settings.
(
→
Chapter 2 “2.4 Refocusing”)
Содержание eclipse Ci-L
Страница 1: ...Upright Microscope Instruction Manual M568 E 11 8 NF 1 1 2 M568EN01 ...
Страница 2: ......
Страница 44: ...Chapter 1 Microscopy Procedures 32 Chapter 1 5 Microscopy Procedures Epi fluorescence Microscopy ...
Страница 86: ...Chapter 4 Troubleshooting 74 Chapter 4 Troubleshooting ...
Страница 90: ...Chapter 5 Maintenance and Storage 78 Chapter 5 Maintenance and Storage ...
Страница 94: ...Chapter 6 Specifications and Safety Standards 82 Chapter 6 Specifications and Safety Standards ...