![Nikon eclipse Ci-L Скачать руководство пользователя страница 35](http://html1.mh-extra.com/html/nikon/eclipse-ci-l/eclipse-ci-l_instruction-manual_1691821035.webp)
Chapter 1
Microscopy Procedures
23
Micr
oscopy
Pr
oced
ures
Sensitive T
int
Plate
Mi
croscopy
Chapter 1-4
5
Place a specimen on the stage, and move the
stage to bring the target into view.
6
Remove the analyzer and the polarizer from the
optical path.
Pull out the analyzer IN/OUT knob from the sensitive tint
plate analyzer unit to remove the analyzer from the optical
path.
The polarizer unit for first-order red compensation has not
been installed at this point.
Removing the analyzer from the optical path
7
Focus on the specimen.
(
→
See Chapter 2 “2 Focusing on the Specimen (Vertical Stage Movement)” for
details)
8
Adjust the diopter.
(
→
See Chapter 2, “4 Adjusting the Diopter” for details)
9
Adjust the interpupillary distance.
10
Focus and center the condenser.
(
→
See Chapter 2, “5 Focusing and Centering the Condenser” for details)
Содержание eclipse Ci-L
Страница 1: ...Upright Microscope Instruction Manual M568 E 11 8 NF 1 1 2 M568EN01 ...
Страница 2: ......
Страница 44: ...Chapter 1 Microscopy Procedures 32 Chapter 1 5 Microscopy Procedures Epi fluorescence Microscopy ...
Страница 86: ...Chapter 4 Troubleshooting 74 Chapter 4 Troubleshooting ...
Страница 90: ...Chapter 5 Maintenance and Storage 78 Chapter 5 Maintenance and Storage ...
Страница 94: ...Chapter 6 Specifications and Safety Standards 82 Chapter 6 Specifications and Safety Standards ...