Chapter 1
Microscopy Procedures
25
Micr
oscopy
Pr
oced
ures
Sensitive T
int
Plate
Mi
croscopy
Chapter 1-4
13
Adjust the orientation of the analyzer and the
polarizer.
(1) Fully open the aperture diaphragm.
(2) Pull out one eyepiece from the tube.
(3) Rotate the lambda plate rotation lever (which makes the upper polarizer move around the pivot) to remove the
lambda plate from the optical path. When doing this, support the polarizer unit for first-order red compensation
with hands so that it does not move.
(4) Look into the eyepiece sleeve and rotate the whole polarizer unit until you can identify a dark cross.
(You will see black stripes that change shape as you rotate the polarizer unit.)
(5) Tighten the fixing screw of the polarizer unit to fix the polarizer.
(6) Put the eyepiece back to the tube and set the lambda plate back into the optical path.
(7) Rotate the lambda plate rotation lever to the left/right
limit to make sure that the field of view is colored
magenta on both sides.
Dark cross (crossed Nicols)
You will see a dark cross when the orientation of
the analyzer orthogonally crosses that of the
polarizer.
Dark cross
14
Bring an arbitrary objective into the optical path.
Turn the nosepiece to bring the desired objective into the optical path.
Swing-out condenser 1-100x
Mounting a 1-100x swing-out condenser on the
Ci-L main body when using the 1x objective may
result in uneven illumination around the field of
view.
15
Focus on the specimen.
(1) Look into the eyepiece, and adjust the brightness of the field of view by turning the dia-illumination brightness
control knob. You can also adjust the brightness with an ND filter for Ci-S.
(2) Rotate the stage knob to move the stage and bring the target into the optical path.
(3) If the specimen is not in focus, turn the focus knob to focus on it.
16
Adjust the field diaphragm.
Turn the field diaphragm dial to adjust the field diaphragm
so that it almost circumscribes the field of view.
Size of the field diaphragm
Normally, adjust the field diaphragm so that it
almost circumscribes the field of view. Opening
the field diaphragm too much results in stray light
entering the field of view, generating flare and
reducing the image contrast. In addition, the
sample will become decolorized over a wider
area.
Field diaphragm's adjustment timing
Be sure to adjust the field diaphragm each time
you change the objective.
Circumscribe around the field of view
Adjusting the field diaphragm
Field of view
Field diaphragm
Field
diaphragm
dial
Содержание eclipse Ci-L
Страница 1: ...Upright Microscope Instruction Manual M568 E 11 8 NF 1 1 2 M568EN01 ...
Страница 2: ......
Страница 44: ...Chapter 1 Microscopy Procedures 32 Chapter 1 5 Microscopy Procedures Epi fluorescence Microscopy ...
Страница 86: ...Chapter 4 Troubleshooting 74 Chapter 4 Troubleshooting ...
Страница 90: ...Chapter 5 Maintenance and Storage 78 Chapter 5 Maintenance and Storage ...
Страница 94: ...Chapter 6 Specifications and Safety Standards 82 Chapter 6 Specifications and Safety Standards ...