7
DATA FORMAT
:
Format of Entire Data
2
The LM transmits the following types of data. These types of data are output according to the lens
measurement mode, whether the data has been measured or not, and the condition for the lenses
(single, left or right).
{
: Output if measured ×: Not output
*1: LM-1200 only.
When the parameter is not set to “ON”, measurement or data output is not performed.
*2: 1st addition power only
*3: In the case of “Single (not the right- nor left-eye lens)” state, only single data is output.
When both left- and right-eye lenses are measured, the data is output in following order:
“Right” “Left”.
*4: In the prism layout, the measured value is output for the PRISM value.
Measured lens mode
Lens
*3
Single
vision
Progressive
power
Contact
lens
Prism
layout
Single
Right
Left
1) Manufacturer/model name
{
{
{
{
Once at the beginning
2) SPH, CYL, AXIS
{
{
{
{
{
{
{
3) SE (Spherical Equivalent value)
×
×
{
×
{
{
{
4) ADD (1st/2nd addition power)
{
{
*2
×
×
{
{
{
5) PRISM
{
{
{
{
{
{
{
6) Progressive Length
*1
×
{
×
×
{
{
{
7) Channel Width
*1
×
{
×
×
{
{
{
8) PD
*1
{
{
×
×
×
{
{
9) Near Portion Inside Amount
*1
×
{
×
×
×
{
{
10) Net Prism
*1
{
×
×
×
×
One for both sides
LM-1000/LM-1000P (In the order of 1)
→
Right 2) - 5) data
→
Left 2) - 5) data)
LM-1200
(In the order of 1)
→
Right 2) - 7) data
→
Left 2) - 7) data
→
(8) - 10))
• Measured data has been rounded off according to the cylinder mode, the indication form of
the measured prism data, and indication step specified on the LM.