9
DATA FORMAT
:
Format of each data
2
1) In the case of single data (SPH= +1.00 D, CYL= 0.00 D, AXIS= 0º)
2) In the case of right data (SPH= -11.25 D, CYL= -9.75 D, AXIS= 90º)
3) In the case of left data (SPH= 0.00 D, CYL= +1.50 D, AXIS= 180º)
(3) SE (Spherical Equivalent value, only in the contact lens measuring mode)
The measured SE value is output.
Measured data is output by fixed four digits including a decimal point with a polarity sign and
the leading digit 0 remaining.
1) In the case of single data (SE = +2.00 D)
2) In the case of right data (SE = -10.00 D)
+
0
1
.
0
0
+
0
0
.
0
0
0
0
0
ETB CR
20H 20H 2BH 30H 31H 2EH 30H 30H 2BH 30H 30H 2EH 30H 30H 30H 30H 30H 17H 0DH
SPH (6 characters)
CYL (6 characters)
AXIS
(3 characters)
R
-
1
1
.
2
5
-
0
9
.
7
5
0
9
0
ETB CR
20H 52H 2DH 31H 31H 2EH 32H 35H 2DH 30H 39H 2EH 37H 35H 30H 39H 30H 17H 0DH
SPH (6 characters)
CYL (6 characters)
AXIS
(3 characters)
L
+
0
0
.
0
0
+
0
1
.
5
0
1
8
0
ETB CR
20H 4CH 2BH 30H 30H 2EH 30H 30H 2BH 30H 31H 2EH 35H 30H 31H 38H 30H 17H 0DH
SPH (6 characters)
CYL (6 characters)
AXIS
(3 characters)
Identification
code
Single
“S ”(S, Space)
Right
“SR”
Left
“SL”
S
+
0
2
.
0
0
ETB CR
53H 20H 2BH 30H 32H 2EH 30H 30H 17H 0DH
SE (6 characters)
S
R
-
1
0
.
0
0
ETB CR
53H 52H 2DH 31H 30H 2EH 30H 30H 17H 0DH
SE (6 characters)
Содержание LM-1000