21
DATA FORMAT
:
Example of Data Output
2
(7) Sample 2 of left and right data in normal measurement mode with PD measurement (LM-
1200)
Measurement conditions:
Single vision lens measurement with the left-eye lens designated
Right: SPH = -1.25 D, CYL = 0.75 D, AXIS = 120º
No prism measurement performed
RPD = 31.5 mm
Left: SPH = -2.00 D, CYL = -0.50 D, AXIS = 180º
No prism measurement performed
No LPD measurement
Total PD is undecided.
CR code = Added
• No PD outputs because LPD has not been measured.
• Net PRISM measurements will also be excluded from the output unless prism has been
measured for both sides.
SOH
D
L
M
STX
→
To be continued
01H
44H 4CH 4DH
02H
I
D
N
I
D
E
K
/
L
M
-
1
2
0
0
ETB CR
→
To be
continued
49H 44H 4EH 49H 44H 45H 4BH 2FH 4CH 4DH 2DH 31H 32H 30H 30H 17H 0DH
Company name/ Model name
R
-
0
1
.
2
5
-
0
0
.
7
5
1
2
0
ETB CR
→
To be
continued
20H 52H 2DH 30H 31H 2EH 32H 35H 2DH 30H 30H 2EH 37H 35H 31H 32H 30H 17H 0DH
Right SPH
Right CYL
Right AXIS
L
-
0
2
.
0
0
-
0
0
.
5
0
1
8
0
ETB CR
→
To be
continued
20H 4CH 2DH 30H 32H 2EH 30H 30H 2DH 30H 30H 2EH 35H 30H 31H 38H 30H 17H 0DH
Left SPH
Left CYL
Left AXIS
EOT
0
B
6
A
CR
04H
30H 42H 36H 41H 0DH
Checksum