18
4.5 Function mode
Turn the power ON while pressing the “MODE” to enter the function mode.
During the function mode, turn the dial (CW/CCW) to change the settings.
And press “MODE” to save the setting and move to the next setting item. (F1
⇒
F2
⇒
F3
⇒
F4
⇒
F5
⇒
F6 end )
If the power is turned OFF in the middle of the function mode, the changed value will not be saved in the memory.
4.5.1 The setting items
The setting items in the following table can be configured in the function mode.
F1
Set to add decimal point (for INT and EXT FPM mode)
Add decimal point
No decimal point
F2
Trigger edge setting (for EXT)
Flash in the rising edge.
Flash in the falling edge.
F3
Auto emission stop time setting
Emission is performed continuously
Stop emission automatically when no operation is performed for a certain period of time
and display “OFF”.
The available setting time is within the range between 0 [min] and 120 [min],
which can be changed by 1 [ms]
F4
Choose the input circuit
Open collector input
Voltage pulse input
F5
Measurement range setting (for INT)
The maximum value of the emission count can be set within the range between 30 and 3000 [fpm]
(The value is limited to the maximum.)
F6
Set the pulse width of the synchronous output signal to 400 [µs] or 150 [µs]
The pulse width is 400 [µs]
The pulse width is 150 [µs]
Содержание DT-3015P
Страница 36: ...35 5 2Externaldimensions...
Страница 38: ......
Страница 39: ......
Страница 40: ...Address 1 Terada Kohtari Nagaokakyo city Kyoto 617 0833 Japan Phone 81 75 958 3621...