27
Adjusting the sensitivity of the optical system
* Because this adjustment needs to be made when the system is in normal condition, ensure
to conduct overhaul work (cleaning or replacing the primary side line of the optical system)
in advance.
1)
After one hour has passed since the system is turned on, open the Maintenance screen.
Then select [Parameter 2] tab and click [Optical Adjust] in AFS settings.
2)
Click [SPAN] in Ref settings to automatically adjust the voltage of the Ref detector to
4.0V.