15
(3) Error message 2
Potential errors when the TC-WA is connected
Error
code
Message
Description and corrective action
11 HEATER BROKEN (H3)
The temperature of H3 does not
rise enough.
Check whether the TC-WA is
turned on.
Generated if the temperature of H3 is below 400°C
on completion of the transfer.
Check the power cable and switch of the TC-WA. If
there is no problem with them, contact us.
12 TC-WA TIME OUT
ERROR (TABLE)
The operation of the table motor
cannot be completed.
A failure has occurred while the table is actuated.
Turn off the power, open the front cover, and check
the status.
If the problem cannot be corrected, contact us.
13 TC-WA TIME OUT ERROR (LIFT)
The operation of the lift motor
cannot be completed.
A failure has occurred while the lift is actuated.
Turn off the power, open the front cover, and check
the status.
If the problem cannot be corrected, contact us.
14 TC-WA LIFT UP ERROR
The lift does not move up.
Generated if the lift does not move up during
loading of the collector tube into the heating
section.
Turn off the power, open the front cover, and check
the status.
If the problem cannot be corrected, contact us.
15 TC-WA NO SAMPLE
The measured collector tube does
not move down.
Generated if the measured collector tube does not
move down.
Turn off the power, open the front cover, and check
the status.
If the problem cannot be corrected, contact us.
16 TC-WA MOTOR ALARM (TABLE)
An alarm is generated from the
motor driver IC for table.
Generated if a failure occurs in the motor driver IC
for table.
Restart the system. If the error still persists, contact
us.
17 TC-WA MOTOR ALARM (LIFT)
An alarm is generated from the
motor driver IC for lift.
Generated if a failure occurs in the motor driver IC
for lift.
Restart the system. If the error still persists, contact
us.
When the error 11 is detected, the system bypasses the subsequent sequences and
proceeds to the cooling process. When H3 has cooled down, the system stops.
When the errors from 12 through 17 are detected, the system immediately cancels the
measurement and stops.
(4) Error message 3
Potential errors when the gas meter is connected
Error
code
Message
Description and corrective action
19 GAS METER ERROR
Generated if the sampling pulse from the gas
meter cannot be detected during measurement in
the direct sampling mode.
When the error is detected, the system bypasses the subsequent sequences and stops.