26
Heater H2
1)
Remove the 2 screws of the re-collecting furnace (H2) and remove the cover.
2)
Remove the reducing Teflon joint and replace the heater and/or thermocouple.
When you fix the thermocouple in place, ensure that its tip is in contact with the heater.
Gas scrubbing bottle and dehumidifying bottle
For how to detach and attach the gas scrubbing bottle and dehumidifying bottle, refer to 3.4.3
“Cleaning the gas scrubbing bottle and dehumidifying bottle.”
Cell