7 DIAGNOSIS
7.1 Displaying the diagnoses
The diagnoses are identified by a
DXXX numbering on the bottom section of
the screen.
Please refer to the table on the next
page to validate the diagnosis, identify the
cause and treatment.
7.2 Cancelling the D001 to D004 and
D007 diagnoses
These diagnoses are related to the pH
part of your STERILOR Duo.
Before cancelling a diagnosis, always
check what caused it and correct it by
following the instructions in this manual.
Press simultaneously on the keys 7 and
10:
•
The display is set on the pH value analysed,
•
The diagnosis display switches off.
You do not need to calibrate the analysis
electrode after this operation which does
not change the device's internal
parameters.
7.3 Cancelling the D005 and D006
diagnoses
These diagnoses are related to the pH
part of your STERILOR Duo.
Before cancelling a diagnosis, always
check what caused it and correct it by
following the instructions in this manual.
1.
Switch off the device with the switch
located on the right-hand side, wait for 5
seconds.
2.
Switch it back on by pressing on the buttons
7 and 10 at the same time.
3.
Hold down both buttons with your fingers
until RST (reset) is displayed instead of pH,
the error must disappear.
4.
Switch off the device again and switch it
back on to make sure that the diagnosis is
no longer displayed, wait for 5 seconds.
5.
If it reappears, repeat the operation
pressing the buttons more firmly.
Once this signal has been cancelled, the
electrode must be calibrated
Indeed, this operation reinitialises all of
the data stored in the device's memory, and
in particular the data concerning the status
of the analysis electrode.
Содержание STERILOR Duo