3. Attach the Nexalin Patient Cable to the electrode (Section 4 - Attaching the Nexalin
Patient Cable). Wait approximately 5 to 10 seconds for the light to stop flashing.
4. Once the CHECK ELECTRODE LED is off, resume the session by pressing the
CURRENT ON button.
5. The CURRENT LEVEL will ramp up and the Nexalin ADI Therapy session will continue.
C. If the Nexalin Patient Cable & Electrode Clips are properly attached and no electrodes have fallen
off, gently run your fingers over each electrode to ensure there are no wrinkles and
the electrodes are properly adhering to the patient’s skin. Wait approximately 5 to 10
seconds for the light to stop flashing. Once the CHECK ELECTRODE LED is off, resume
the session by pressing the CURRENT ON button. The CURRENT LEVEL will ramp up and the Nexalin ADI
Therapy session will continue.
Battery LED
In case of a power failure or a blown fuse, the Nexalin ADI Device has an internal backup battery
designed to supply uninterrupted power to a Nexalin ADI Therapy session already in progress. If the
internal backup battery of the Nexalin ADI Device is in use, the amber BATTERY LED will light up or
flash, depending upon the situation. Do not initiate a new Nexalin ADI Therapy session with a Nexalin
ADI Device displaying a lit or flashing BATTERY LED.
BATTERY LED is on but not flashing — The device is operating on the backup battery.
BATTERY LED is flashing — The device is operating on the backup battery, and the backup battery level
is low or has become low during a battery-powered Nexalin ADI Therapy session.
1. Verify the power source or, if used, the universal power supply has power supplied to it.
2. If there is no power failure, verify that the power cord is plugged in properly at both the power
source and the Nexalin ADI Device. If a universal power supply or alternate power source is being used,
verify that it is connected properly.
3. If the BATTERY LED is still lit or flashing, refer to Section 5 - Replacing the Nexalin ADI De- vice
Fuses, or contact the Nexalin Support Center for assistance.
CAUTION!
The internal backup battery is not designed to operate for more than 40 minutes. Never initiate a new
Nexalin ADI Therapy session on backup battery power.