8.3. DISPERSED RADIATION
Measurements of dispersed radiation are significantly affected by environmental conditions, such as the composition
of walls and their position, therefore, in certain circumstances, values can be significantly different.
The measurement points used are at a distance of 0.5 m, 1.0 m, 2.0 m and 3.0 m from a central axis.
Circular measurement curves refer to the above mentioned axis.
Dispersed radiation is measured in standard panoramic X-ray mode or in 3D mode with medium field in patients with
a large build, selected with a phantom cylinder PMMA (Ф 16 cm diameter x 17 cm height) to simulate the patient’s
head.
1) Standard panoramic X-ray: Dispersed radiation measured at the maximum percentage of use permitted by the X-
ray generator (it corresponds to an average anode power of 42 W).
Distance between the rotation axis and the
measurement point
(Circular measurement points)
Dispersed radiation*
0.5 m
60 µGy / h
1.0 m
15 µGy / h
2.0 m
4 µGy / h
3.0 m
1.5 µGy / h
2) Standard panoramic X-ray: Dispersed radiation at the average percentage of use or 4 examinations per hour.
Distance between the rotation axis and the
measurement point
(Circular measurement points)
Dispersed radiation*
0.5 m
12 µGy / h
1.0 m
3.0 µGy / h
2.0 m
0.8 µGy / h
3.0 m
0.3 µGy / h
3) 3D 10x8 standard/regular: Dispersed radiation measured at the maximum percentage of use permitted by the X-
ray generator (it corresponds to an average anode power of 42 W).
Distance between the rotation axis and the
measurement point
(Circular measurement points)
Dispersed radiation*
0.5 m
670 µGy / h
1.0 m
180 µGy / h
2.0 m
45 µGy / h
3.0 m
21 µGy / h
4) 3D 10x6 standard/regular: Dispersed radiation at the average percentage of use or 4 examinations per hour.
Distance between the rotation axis and the
measurement point
(Circular measurement points)
Dispersed radiation*
0.5 m
45 µGy / h
1.0 m
12 µGy / h
2.0 m
3
µGy / h
3.0 m
1.4 µGy / h
*
It is the highest value at 15 cm over the horizontal intersection plane with the patient’s bite. The other values of
the vertical axis are lower than these values.
Содержание GiANO HR
Страница 1: ...97050977 Rev 02 2018 11 NewTom GiANO HR EN ...
Страница 2: ...2 OPERATOR S MANUAL EN ITALIANO ...
Страница 75: ...EN OPERATOR S MANUAL 75 Characteristics of X ray tubes CANON TOSHIBA D 0510SB ...
Страница 76: ...76 OPERATOR S MANUAL EN CEI OPX105 12 ...
Страница 78: ...78 OPERATOR S MANUAL EN 8 4 ISODOSE CURVES FOR CBCT EXAMINATIONS 8 5 ISODOSE CURVES FOR PAN EXAMINATIONS ...
Страница 88: ...Figure 1 Figure 2 ...