OPERATOR'S MANUAL
45
5.4. PATIENT POSITIONING
Make the patient access the machine and set any cranium positioning only when the machine has stopped moving
and the console is in the patient positioning page.
If you want to stop the machine movement, press the CONFIRMATION button again.
Here you find instructions on how to
position the patient using the
positioning devices provided for.
It is shown how to position the
patient with the help of laser traces.
The height of the column (A) and the
position of the Vertical focusing laser
(B) are also indicated. Use the
“arrow” keys to move the column or
the vertical laser position by 1 mm
steps, in the direction indicated by
the arrow.
Lasers can be turned on and off by using the switch (C).
Once the wanted position has been reached, touch NEXT to go to the EXECUTION OF EXAMINATION screen.
When the patient accesses the machine during the “Patient Positioning” or the “Execution of Examination”
(and anyway before starting the examination), make sure that the X-ray device is not accidentally hit: in this
case, it is recommended to make the patient exit the machine and reposition the X-ray device, returning to
the “Examination Settings” screen and repeating the procedure.
5.4.1. LASER TRACES
The X-ray device features four light traces which help positioning the patient:
1) Upper Horizontal Laser Trace
used in all PAN, DENT, SENI,
TMJ, 3D examinations
2) Sagittal Vertical Laser Trace
used in all the examinations
3) Vertical Focusing Laser Trace
used in PAN, DENT, SENI, TMJ
examinations
4) Lower Horizontal Laser Trace
used in 3D, CEPH examinations
Upper horizontal trace (1)
it is generated by a laser projector placed on the X-ray generator side; it can be moved upward or downward to be
adapted to different dimensions, using the suitable lever beside the light opening. This trace is used in PAN and
DENT examinations to ensure
that the patient’s Frankfurt plane is horizontal and, more generally, to ensure correct
positioning.
Содержание GiANO HR
Страница 1: ...97050977 Rev 02 2018 11 NewTom GiANO HR EN ...
Страница 2: ...2 OPERATOR S MANUAL EN ITALIANO ...
Страница 75: ...EN OPERATOR S MANUAL 75 Characteristics of X ray tubes CANON TOSHIBA D 0510SB ...
Страница 76: ...76 OPERATOR S MANUAL EN CEI OPX105 12 ...
Страница 78: ...78 OPERATOR S MANUAL EN 8 4 ISODOSE CURVES FOR CBCT EXAMINATIONS 8 5 ISODOSE CURVES FOR PAN EXAMINATIONS ...
Страница 88: ...Figure 1 Figure 2 ...