90
DH
Description
Home position set.
Syntax
xx
DH
nn
Remarks
This command is used to define the “home” position for an axis. The home position is set to 0 if
this command is issued without “nn” value. Upon receipt of this command, the controller will set
the present position to the specified home position. The move to absolute position command
(PA) uses the “home” position as reference point for moves.
Argument
Value
Description
xx
1 to 2
Axis number
nn
Integer
Home position (steps). Default = 0 steps. Valid values
between
-2147483648
and
+2147483647
.
Example
1DH
(set motor 1 counter to 0)
2DH1000
(set motor 2 counter equal to 1000)
See Also
DH?, TP?, PA, OR, MT, MZ
DH?
Description
Home position query.
Syntax
xx
DH
?
Remarks
This command is used to query the home position value for an axis.
Argument
Value
Description
xx
1 to 2
Axis number
Response
Value
Description
Value
Integer
Home position setting (steps)
Example
2DH?
(get home position)
0
(reply; home position is 0)
See Also
DH, TP?
Содержание New Focus Picomotor 8743-CL
Страница 1: ...Picomotor Controller Driver User s Manual Model 8743 CL...
Страница 10: ...x Preface...
Страница 12: ...xii Preface Device Address Window 70 Internet Browser View 71 Command Syntax Diagram 73...
Страница 40: ...40 Multiple Units Stacking Interconnection Multiple Units Stacking Interconnection 8742 controllers shown...
Страница 59: ...59 Cycle Tab Cycle Motion...
Страница 60: ...60 Terminal Tab Command Line Terminal...
Страница 64: ...64 Options Window Connection Options Setup...
Страница 65: ...65 Properties Window USB and Ethernet Properties...
Страница 66: ...66 Controller Scan Scan RS 485 LAN...
Страница 67: ...67 Address Conflict Resolution Window Address Conflict Resolution...
Страница 70: ...70 Device Address Window...