88
DB
Description
Deadband set.
Syntax
xx
DB
nn
Remarks
This command is used to set the position deadband value for an axis. Since a majority of
electro-mechanical systems have mechanical backlash or frictional hysteresis, closed-loop
positioning can at times lead to oscillation or limit cycling of the systems around a desired
position. In such situations, setting position deadband value judiciously can avoid limit cycling
of the systems.
Note that this command is effective only during position regulation (holding position) as
opposed to while moving. Also, the positioning units must be in ‘Counts’ (see SN command)
and closed-loop positioning must be enabled (see MM command) for DB command to take
effect.
NOTE: An error will be generated if the controller makes more than 100 consecutive
unsuccessful attempts to correct the stage position, but is unable to bring the
following error to within specified deadband setting.
Argument
Value
Description
xx
1 to 2
Axis number
nn
0 to
2147483647
Deadband value (units). Default = 0
Example
2DB0
(set deadband value to 0 encoder counts)
2SN1
(set units to encoder counts)
2FE1000
(set following error threshold to 1000 counts)
2CL0.1
(set closed-loop update interval to 0.1 second)
SM
(save settings to non-volatile memory)
2MM1
(enable closed-loop positioning)
2PA+100
(move to encoder count po100)
See Also
DB?, SN, MM, FE, CL
Содержание New Focus Picomotor 8743-CL
Страница 1: ...Picomotor Controller Driver User s Manual Model 8743 CL...
Страница 10: ...x Preface...
Страница 12: ...xii Preface Device Address Window 70 Internet Browser View 71 Command Syntax Diagram 73...
Страница 40: ...40 Multiple Units Stacking Interconnection Multiple Units Stacking Interconnection 8742 controllers shown...
Страница 59: ...59 Cycle Tab Cycle Motion...
Страница 60: ...60 Terminal Tab Command Line Terminal...
Страница 64: ...64 Options Window Connection Options Setup...
Страница 65: ...65 Properties Window USB and Ethernet Properties...
Страница 66: ...66 Controller Scan Scan RS 485 LAN...
Страница 67: ...67 Address Conflict Resolution Window Address Conflict Resolution...
Страница 70: ...70 Device Address Window...