Description
Velocity set
Syntax
xx
VA
nn
Remarks
This command is used to set the velocity value for an axis. The velocity setting specified will
not have any effect on a move that is already in progress. If this command is issued when an
axis’ motion is in progress, the controller will accept the new value but it will use it for
subsequent moves only. The maximum velocity for a ‘Standard’ Picomotor is 2000 steps/sec,
while the same for a ‘Tiny’ Picomotor is 1750 steps/sec.
Argument
Value
Description
xx
1 to 2
Axis number
nn
1 to 2000
Velocity (steps/sec). Default = 2000 steps/sec
Example
1VA1750
(Set motor 1 velocity to 1750 steps/sec)
See Also
VA?, AC, AC?
VA?
Description
Velocity query.
Syntax
xx
VA
?
Remarks
This command is used to query the velocity value for an axis.
Argument
Value
Description
xx
1 to 2
Axis number
Response
Value
Description
Value
Integer
Velocity setting (steps/sec)
Example
1VA?
(Get motor 1 velocity)
1750
(Returns 1750; motor 1 velocity)
See Also
VA, AC, AC?
Содержание New Focus Picomotor 8743-CL
Страница 1: ...Picomotor Controller Driver User s Manual Model 8743 CL...
Страница 10: ...x Preface...
Страница 12: ...xii Preface Device Address Window 70 Internet Browser View 71 Command Syntax Diagram 73...
Страница 40: ...40 Multiple Units Stacking Interconnection Multiple Units Stacking Interconnection 8742 controllers shown...
Страница 59: ...59 Cycle Tab Cycle Motion...
Страница 60: ...60 Terminal Tab Command Line Terminal...
Страница 64: ...64 Options Window Connection Options Setup...
Страница 65: ...65 Properties Window USB and Ethernet Properties...
Страница 66: ...66 Controller Scan Scan RS 485 LAN...
Страница 67: ...67 Address Conflict Resolution Window Address Conflict Resolution...
Страница 70: ...70 Device Address Window...