83
AC
Description
Acceleration set.
Syntax
xx
AC
nn
Remarks
This command is used to set the acceleration value for an axis. The acceleration setting
specified will not have any effect on a move that is already in progress. If this command is
issued when an axis’ motion is in progress, the controller will accept the new value but it will
use it for subsequent moves only.
Argument
Value
Description
xx
1 to 2
Axis number
nn
1 to 200000
Acceleration (steps/sec
2
). Default = 100000 steps/sec
2
Example
2AC150000
(set acceleration to 150,000 steps/sec
2
)
See Also
AC?, VA, VA?
AC?
Description
Acceleration query.
Syntax
xx
AC
?
Remarks
This command is used to query the acceleration value for an axis.
Argument
Value
Description
xx
1 to 2
Axis number
Response
Value
Description
Value
Integer
Acceleration setting (steps/sec
2
)
Example
2AC?
(get acceleration setting)
150000
(reply; 150,000 steps/sec
2
)
See Also
AC, VA, VA?
Содержание New Focus Picomotor 8743-CL
Страница 1: ...Picomotor Controller Driver User s Manual Model 8743 CL...
Страница 10: ...x Preface...
Страница 12: ...xii Preface Device Address Window 70 Internet Browser View 71 Command Syntax Diagram 73...
Страница 40: ...40 Multiple Units Stacking Interconnection Multiple Units Stacking Interconnection 8742 controllers shown...
Страница 59: ...59 Cycle Tab Cycle Motion...
Страница 60: ...60 Terminal Tab Command Line Terminal...
Страница 64: ...64 Options Window Connection Options Setup...
Страница 65: ...65 Properties Window USB and Ethernet Properties...
Страница 66: ...66 Controller Scan Scan RS 485 LAN...
Страница 67: ...67 Address Conflict Resolution Window Address Conflict Resolution...
Страница 70: ...70 Device Address Window...