24
Settings
3-4. Settings
From the Setup Menu screen, it is possible to change the target gas; convert to other
gas’s concentration; turn on/off Bluetooth; set the time span for trend graph display;
set logging interval rate; turn on/off logging; delete logs; adjust the audio volume; set
the alarm set value; perform an alarm test; adjust the clock; and select the language.
*1. The target gas can be changed (selected) if the unit is a model capable of detecting more
than one combustible gas.
*2. The mode is enabled only with XP-3310II, XP-3318II, XP-3360II-W, and XP-3368II-W
units, in which GAS1 (target gas) is set to methane, isobutane, or propane. See
“Convertible Gas List” on page 27 for the 32 convertible gas options.
*3. It is possible to view log data as well as real-time measurement data, etc., in a graph form
by using an Android (ver. 7.0 or higher) or iOS (ver. 10.0 or higher) device with built-in
Bluetooth. To do this, download a free app “Utility for gas detector” from the Google Play
or App store.
It is also possible to read out log data from the detector in a graph form/CSV format, by
using a PC (Windows 8.1 and 10) and a data logger software (XP-3000IIL, separately
sold). For the readout procedure, see the data logger software’s instruction manual.
*4. Settable for the unit with Bluetooth function installed.
Setup item
Function and setting
Ref.
page
Change target gas
Changes (selects) the target gas
*1
26
Convert to other gas’s
concentration
Calculates and displays other gas
concentration (estimation)
*2
26-27
Bluetooth on/off
Sets the Bluetooth to on/off
*4
For connectable devices, see
*3
27
Trend graph span setting
Sets the trend graph’s time span to
display on the LCD
Settable span is 1, 2, 3, 4, or 5 minutes.
28
Logging interval rate setting,
logging on/off, and
log deletion
Sets the logging interval rate
*4
Sets the data logging to on/off
*4
Deletes all the logs
*4
28-30
Audio volume control
Adjusts the audio level (gas alarm
sound, fault alarm sound and button
tone)
31
Alarm set value setting and
alarm test
Sets the alarm set value
Checks an alarm operation (alarm light
and sound) by simulating a gas alarm
condition.
32-33
Clock adjustment
Adjusts the current time
33
Language selection
Selects the language to display,
between Japanese and English
33
Содержание XP-3000II Series