background image

22

 

M1305-0050   Innova CO-170 CO

2

 Incubator   

User’s Guide 

3.2  Preparing for Operation 

 

1.  Remove the black protective cover from the CO

2

 sensor (located on the back 

surface of the internal chamber), and store it for use when you clean the 
incubator.  There is a storage holder for the cover on the front the unit, inside 
the door. 

 
2.  Ensure that the white porous sensor cover remains in place. 
 
3.  Using the power cord provided, connect the incubator to a grounded power supply. 

 

4.  Switch the incubator ON using the on/off switch at the rear of the cabinet.  The 

display will illuminate immediately. 

 

5.  Turn on the CO

2

 gas supply with the pressure regulator set to 5 psi or 0.35 bar.  

NOTE: 

Be certain to check for leaks in the CO

2

 connections to avoid depleting 

your CO

2

 gas supply.  This can be accomplished using a solution of 

soapy water applied to each fitting and checking for bubbles.  If any 
bubbles are noted, readjust the fitting. 

 

6.  The chamber setpoints are pre-programmed at 37.0°C and 5% CO

2

.  Leave the 

incubator on until the programmed chamber temperature and CO

2

 concentration 

has been reached (

please see 

NOTE

 below

). 

NOTE: 

The incubator’s CO

2

 valve is disabled until the incubator reaches the 

temperature setpoint.  After reaching the temperature setpoint, the 
CO

2

 valve is activated allowing the unit to reach the CO

2

 setpoint.  If 

power is interrupted to the incubator long enough for the temperature 
to drop below setpoint, the CO

2

 valve will be deactivated until setpoint 

is again achieved.

  (This serves to avoid spurious CO

2

 readings while 

the incubator is reaching its temperature setpoint) 

 

7.  Leave the incubator running for at least two hours (preferably overnight) to 

allow conditions to stabilize.  

 

3.3  Using the Humidity Tray & Humidity Control 

 

If humidification is required, the humidity tray should be filled with 1.5 liters of 
warm (

 37.0ºC) distilled water at this time.

  

 
 

Содержание Innova CO-170

Страница 1: ...1305 0050 Revision A June 4 2004 NEW BRUNSWICK SCIENTIFIC CO INC BOX 4005 44 TALMADGE ROAD EDISON NJ 08818 4005 Telephone 1 732 287 1200 1 800 631 5417 Fax 732 287 4222 Telex 4753012 NBSCO Internet ht...

Страница 2: ...2 M1305 0050 Innova CO 170 CO2 Incubator User s Guide...

Страница 3: ...1265 Zhongshan Road W Shanghai 200051 P R China Tel 86 21 3223 0203 Fax 86 21 6278 7182 E mail nbschc online sh cn THE NETHERLANDS New Brunswick Scientific BV Kerkenbos 1101 6546 BC Nijmegen P O Box...

Страница 4: ...4 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 5: ...ersonal injury can occur Please read the entire User s Guide before attempting to use this unit Do not use this equipment in a hazardous atmosphere or with hazardous materials for which the equipment...

Страница 6: ...6 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 7: ...e right to change information in this document without notice Updates to information in this document reflect our commitment to continuing product development and improvement Manual Conventions NOTE C...

Страница 8: ...8 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 9: ...d our obligation under this warranty is limited to repairing or replacing the instrument or part thereof which shall within 2 years following date of shipment prove to be defective after our examinati...

Страница 10: ...10 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 11: ...S 25 3 6 1 DATE TIME ADJUST 26 3 6 2 AUDIO VOLUME ADJUST 26 3 6 3 PROGRAMMABLE CO2 AUTOZERO 26 3 6 4 DATALOGGER 27 3 6 5 POWER FREQUENCY 27 3 6 6 DISABLE 27 3 6 7 DECONTAMINATION Model Option 28 3 7 D...

Страница 12: ...OPTIONS ACCESSORIES 45 7 1 OPTIONS 45 7 2 AVAILABLE ACCESSORIES 45 8 DRAWINGS TABLES 47 8 1 LIST OF DRAWINGS 47 8 2 LIST OF TABLES 47 9 APPENDIX A HIGH TEMPERATURE DECONTAMINATION OPTION 49 9 1 USING...

Страница 13: ...excellent reliability and remaining unaffected by humidity The CO2 system has a programmable automatic zero system AutoZero to re reference the sensor baseline to atmospheric CO2 levels at regular int...

Страница 14: ...ependently from the main control system The incubator features multiple options that can be installed to simplify maintenance and provide superior control over experimental conditions For example high...

Страница 15: ...leave it on the pallet You must have at least one person at either side to safely lift the unit Supported by the base only remove the incubator by lifting it from the delivery pallet Locate and remov...

Страница 16: ...Provided There is a holder at the front of the unit inside the door to store the black sensor cover 2 3 Utilities In order to use the incubator you will need Table 2 Utilities Utility Requirement 100...

Страница 17: ...ackable the top panel will NOT support a second unit A second incubator may be safely stacked on top of another identical incubator by using the custom designed stacking stand available as an accessor...

Страница 18: ...R Rear goes to the rear of the chamber as shown in Figure 1a below The suction cups will adhere to the chamber walls even if they are dry but if you feel it is necessary you can dampen them with disti...

Страница 19: ...grip facing forward Make sure the tray is pushed all the way against the rear wall of the chamber Figure 3 Installing Removing the Humidity Tray Connect the CO2 gas supply 1 Connect the unit to the CO...

Страница 20: ...20 M1305 0050 Innova CO 170 CO2 Incubator User s Guide Figure 4 AutoZero HEPA Filter Fitting...

Страница 21: ...is also how the screen should look when you switch the incubator on The purpose of each Function Key is identified at the bottom of the display above each key the function may change from screen to s...

Страница 22: ...to each fitting and checking for bubbles If any bubbles are noted readjust the fitting 6 The chamber setpoints are pre programmed at 37 0 C and 5 CO2 Leave the incubator on until the programmed chambe...

Страница 23: ...t adjustable The internal chamber will reach 95 relative humidity at 37o C using the 1 5 liter humidity tray An optional 2 5 liter tray allows for 97 relative humidity at 37o C NOTE The humidity tray...

Страница 24: ...To set the User Access Code 1 In the PROG screen accessed by pressing the PROG function key the user access code will be displayed as a series of four asterisks 2 Use the left and right direction keys...

Страница 25: ...is running 3 When the countdown is complete the unit is ready to use NOTE It may be necessary to open the door momentarily if after performing an AutoZero the CO2 level is too high 3 6 USER Settings I...

Страница 26: ...references 3 6 3 PROGRAMMABLE CO2 AUTOZERO When you select this feature the PROGRAM CO2 AUTOZERO screen see Figure 7 below allows you to program the AutoZero frequency and time or to run the AutoZero...

Страница 27: ...end that you only change the time setting shortly before you start to use the incubator If the AutoZero function is to be run manually simply press the START function key within the PROGRAM CO2 AUTOZE...

Страница 28: ...lace during the cycle to ensure that everything within the chamber is decontaminated prior to resumption of activity For a full explanation of this feature see Section 9 1 3 7 DATALOGGER Screen The DA...

Страница 29: ...18 seconds while the temperature is outside the specification of 0 1 C and each reading is shown as a single pixel When the temperature has settled within specification the recording is compressed to...

Страница 30: ...cting the port to a PC loaded with InnovaLog Software see Section 7 2 3 7 3 CO2 GRAPH DOOR OPEN BAR CHART These graphs record in a way similar to the Chamber Temperature graphs The specification for C...

Страница 31: ...ture and CO2 High and Low Alarm setpoints automatically adjust to within 0 5 of the temperature and CO2 setpoints The alarm setpoints can also be manually adjusted Figure 11 CHAMBER ALARMS Screen To a...

Страница 32: ...Function and Figure 13 explain in detail how the Chamber Alarms operate 3 8 1 Chamber Alarm System Function When the incubator is switched ON or after the temperature and CO2 levels have been reprogra...

Страница 33: ...are stored in the Datalogger Figure 13 Chamber Alarm System Flow Chart 15 MINUTES ALARM DELAY TIMEOUT COMPLETED ARE TEMPERATURES AND CO2 LEVELS WITHIN 0 1 OF SET POINT CHANGE PROGRAMMED VALUES OPEN I...

Страница 34: ...he ALARM screen to deal with the situation that triggered the alarm 3 9 DIAGNOSTIC Menu Screen The diagnostics screen Figure 15 below contains technical information regarding the status of many of the...

Страница 35: ...ser selectable categories of abbreviated information found in the user manual All the major systems are covered in the help menu including help on installing the unit If the user manual is misplaced i...

Страница 36: ...36 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 37: ...king conditions a typical large cylinder should last approximately 12 months frequent door openings will deplete the supply more rapidly If there is a significant drop at the cylinder pressure of 725...

Страница 38: ...ing Turn off the CO2 gas by re programming the setpoint for CO2 to 0 0 to prevent CO2 from being injected into the chamber and giving a false reading A flow rate 0 5 liters minute is used to take a sa...

Страница 39: ...d your work area follow these instructions 1 Program 0 0 CO2 and switch off the incubator Unplug the unit from the power supply 2 Dampen a clean cloth with the alcohol solution and wipe down all exter...

Страница 40: ...o be removed from any additional sensors installed such as Oxygen or Humidity 10 Refill the humidity tray as explained in Section 3 3 When you reinstall it ensure that the humidity tray is pushed full...

Страница 41: ...er s Guide 5 3 High Temperature Decontamination If your incubator is supplied with the high temperature decontamination option follow the guidelines outlined in the Appendix Section 9 for correctly an...

Страница 42: ...42 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 43: ...etter than 0 7 minute Gas Connections 1 4 inch or 6mm tubing Required Gas Pressure 5 psi 0 35 bar Relative Humidity Removable stainless steel humidity tray Reservoir Capacity 1 5 liters Humidity Contr...

Страница 44: ...Innova CO 170 Incubator Specifications Grounded Electrical Supply Voltage 100 120V 50 60 Hz or 220 240V 50 60 Hz Power 500 watts Standard models 1000 watts High Temperature Decontamination models Ener...

Страница 45: ...P0628 5500 No RS 232 Port P0628 5510 Yes Centronics Printer Port P0628 5520 Yes 3 Inner Doors P0628 5440 No 6 Inner Doors P0628 5450 No Alarm Relay Contacts P0628 5540 Yes Internal Power Receptacle P...

Страница 46: ...46 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 47: ...Screen 25 7 PROGRAM CO2 AUTOZERO Screen 26 8 DATALOGGER Screen 28 9 ALARM EVENTS Screen 29 10 TEMPERATURE GRAPH DOOR OPEN BAR CHART Screen 30 11 CHAMBER ALARM Screen 31 12 Chamber Alarm Message 32 13...

Страница 48: ...48 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 49: ...If the Humidity Tray Warning Option is fitted ensure the sensor is placed on the bottom shelf prior to starting the cycle This is to protect the sensor from the direct heat at the bottom of the incuba...

Страница 50: ...ween 120 150 C particularly in areas that are inaccessible to chemical decontamination Certain areas of the glass door and inner door seal surface temperatures will be somewhat lower than 120 C 6 Afte...

Страница 51: ...riod indicates the temperature fell below 118 0 C during the decontamination phase Y Temperature increase during cool down phase indicates the temperature rose by more than 2 C during the cooling phas...

Страница 52: ...lows 1 Remove the black plastic plug that covers the Oxygen Sensor located on the back of the unit 2 Disconnect the sensor cable by unplugging the electrical connector grasp the white connector body n...

Страница 53: ...tight Do not over tighten as this will distort the Oxygen Sensor and cause it to malfunction 2 Reconnect the sensor cable by plugging in the connector grasp the white connector body not the wire leads...

Страница 54: ...54 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...

Страница 55: ...ystem 32 Chamber Alarm to Check CO2 Supply 34 Checks Daily 37 Monthly 37 Cleaning 39 CO2 Control Specifications 43 Condensation 37 Minimizing 14 Control Panel 21 23 Copyright Notice 7 D Datalogger 27...

Страница 56: ...requency 27 Pressure Setting 22 Programmable CO2 AutoZero 26 Programming 23 R Relative Humidity Specifications 43 Restart Graphic Record 31 S Sensor Cover 15 39 Setpoints Factory set 22 Setup Concerns...

Отзывы: