5. You can clean the humidity tray by rinsing it in sterile water, wiping it down
with the alcohol solution, and then rinsing it again with sterile water.
6. Wipe down the inside of the chamber with the alcohol/water solution, and leave
it to dry completely.
7. Wipe the internal components of the chamber twice with the alcohol/water
solution. Wipe off excess liquid and leave it to dry completely.
8. Reassemble the shelf racks, shelves and humidity tray before switching the
incubator on. Wipe the inner door seal with the alcohol solution, rinse and
leave to dry.
9. Ensure the protective cover(s) are removed from the sensor(s) and replaced in
the holder for safekeeping (protective covers must also be removed from any
additional sensors installed, such as Oxygen or Humidity).
10. Refill the humidity tray (
as explained in Section 3.3
). When you reinstall it,
ensure that the humidity tray is pushed fully back against the rear wall.
11. Leave the incubator on for at least two hours (preferably overnight) to allow
conditions to stabilize.
12. When the incubator has stabilized, carry out an AutoZero and reprogram the
desired CO
2
level. It may be necessary to open the glass door briefly if, after
performing an AutoZero, the CO
2
level is too high.
CAUTION!
It is very important to ensure that no liquid is spilled onto the
white porous CO
2
sensor cover at the rear of the chamber.
Failure to use the protective cover(s) could result in damage to
the sensor(s) and may affect your warranty.
CAUTION!
Never
use any of the following substances to clean the stainless
steel, or damage will result: Sodium Azide, Aqua Regia, Iodine,
Ferric Chloride or Sulphuric Acid.
Содержание Innova CO-170
Страница 2: ...2 M1305 0050 Innova CO 170 CO2 Incubator User s Guide...
Страница 4: ...4 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Страница 6: ...6 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Страница 8: ...8 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Страница 10: ...10 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Страница 20: ...20 M1305 0050 Innova CO 170 CO2 Incubator User s Guide Figure 4 AutoZero HEPA Filter Fitting...
Страница 36: ...36 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Страница 42: ...42 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Страница 46: ...46 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Страница 48: ...48 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...
Страница 54: ...54 M1305 0050 Innova CO 170 CO2 Incubator User s Guide THIS PAGE IS INTENTIONALLY BLANK...