New Brunswick Scientific Co., Inc.
User’s Guide
111
The four pump control loops can be used to calibrate each pump. To do so, the
control mode and output action of each loop must be changed to
TOTAL
and
DIRECT
. When a loop is configured with a control mode of
TOTAL
, the display
gauge displays the sum of its output since last reset. This control mode is typically
used to monitor accumulated totals of pump or other actuators. Calibrating this
type of loop consists of running the pump or other actuator for a known period of
time, and measuring the accumulated total in volume, mass, etc.
To calibrate a total control loop:
•
In the
GAUGE SCREEN
, set the loop setpoint from the display gauge to zero.
•
From the
GAUGE SCREEN
, select the
CAL BUTTON
. The
CALIBRATION
SELECTION WINDOW
will appear.
•
Select the control loop you wish to calibrate by pressing on it. A
CALIBRATION SCREEN
will appear.
•
In the function window, press the
SET TOTAL FUNCTION
. Enter 0 using the
onscreen keypad and press the
SET TOTAL BUTTON
.
If the pump or actuator used has been calibrated, and its flow or transfer
rate is known:
•
In the function window, select the
RATE FUNCTION
. The
SET TOTAL
BUTTON
will change to a
SET RATE BUTTON
.
Содержание BioFlo 4500
Страница 2: ...BioFlo 4500 M1254 0050 User s Guide ii...
Страница 4: ...BioFlo 4500 M1254 0050 User s Guide iv THIS PAGE INTENTIONALLY BLANK...
Страница 6: ...BioFlo 4500 M1254 0050 User s Guide vi THIS PAGE INTENTIONALLY BLANK...
Страница 8: ...BioFlo 4500 M1254 0050 User s Guide viii THIS PAGE INTENTIONALLY BLANK...
Страница 10: ...BioFlo 4500 M1254 0050 User s Guide x THIS PAGE INTENTIONALLY BLANK...
Страница 12: ...BioFlo 4500 M1254 0050 User s Guide xii THIS PAGE INTENTIONALLY BLANK...
Страница 18: ...BioFlo 4500 M1254 0050 User s Guide xviii...
Страница 34: ...BioFlo 4500 M1254 0050 User s Guide 34 THIS PAGE INTENTIONALLY BLANK...
Страница 42: ...BioFlo 4500 M1254 0050 User s Guide 42 THIS PAGE INTENTIONALLY BLANK...
Страница 60: ...BioFlo 4500 M1254 0050 User s Guide 60 THIS PAGE INTENTIONALLY BLANK...
Страница 82: ...BioFlo 4500 M1254 0050 User s Guide 82 THIS PAGE INTENTIONALLY BLANK...
Страница 88: ...BioFlo 4500 M1254 0050 User s Guide 88 THIS PAGE INTENTIONALLY BLANK...
Страница 114: ...BioFlo 4500 M1254 0050 User s Guide 114 THIS PAGE INTENTIONALLY BLANK...
Страница 138: ...BioFlo 4500 M1254 0050 User s Guide 138 THIS PAGE INTENTIONALLY BLANK...
Страница 144: ...BioFlo 4500 M1254 0050 User s Guide 144 THIS PAGE INTENTIONALLY BLANK...
Страница 152: ...BioFlo 4500 M1254 0050 User s Guide 152 THIS PAGE INTENTIONALLY BLANK...
Страница 154: ...BioFlo 4500 M1254 0050 User s Guide 154...
Страница 160: ...BioFlo 4500 M1254 0050 User s Guide 160 THIS PAGE INTENTIONALLY BLANK...
Страница 162: ...BioFlo 4500 M1254 0050 User s Guide 162 THIS PAGE INTENTIONALLY BLANK...
Страница 164: ...BioFlo 4500 M1254 0050 User s Guide 164 THIS PAGE INTENTIONALLY BLANK...
Страница 168: ...BioFlo 4500 M1254 0050 User s Guide 168 THIS PAGE INTENTIONALLY BLANK...
Страница 186: ...BioFlo 4500 M1254 0050 User s Guide 186 THIS PAGE INTENTIONALLY BLANK...
Страница 194: ...BioFlo 4500 M1254 0050 User s Guide 194 THIS PAGE INTENTIONALLY BLANK...
Страница 196: ...BioFlo 4500 M1254 0050 User s Guide 196 THIS PAGE INTENTIONALLY BLANK...
Страница 208: ...BioFlo 4500 M1254 0050 User s Guide 208 THIS PAGE INTENTIONALLY BLANK...
Страница 212: ...BioFlo 4500 M1254 0050 User s Guide 212 Figure 47 Basic Control Schematics Detail B SEE DETAIL A...
Страница 214: ...BioFlo 4500 M1254 0050 User s Guide 214 Figure 49 Power Schematics Detail A...