NEUROTHERM RADIO FREQUENCY LESION GENERATOR
8-5
MODEL NT1100
®
OPERATORS MANUAL
Document 109.00
10September2012 Issue 06
Level 3 - User warning
Message Scenario
Message Text
Message
code
When trying to print, but there is an error with the
printer.
E.g. out of paper / out of ink / paper jam
Printing error - check printer
Fix problem, press <retry> or press <cancel> to
abort the print.
3-001
When trying to switch into sensory/motor stim with
the stimulation pot switched on
Turn off stimulation control before continuing
3-003
When trying to switch into lesion/pulse mode with
the RF power control switched on
Turn off RF power control before continuing
3-004
When the accept button is pressed in the setup
confirmation screen with No thermocouple elect
selected
No thermocouple elect selected.
Be aware that there will be no temperature
monitoring during the procedure.
3-005
Open circuit condition on Probe 1 with stimulate
control turned on
Unable to stimulate probe 1
Impedance probe 1 open circuit.
Turn off stimulate control and check probe and
reference plate.
3-006
Open circuit condition on Probe 2 with stimulate
control turned on
Unable to stimulate probe 2
Impedance probe 2 open circuit.
Turn off stimulate control and check probe and
reference plate.
3-007
Open circuit condition on Probe 3 with stimulate
control turned on
Unable to stimulate probe 3
Impedance probe 3 open circuit.
Turn off stimulate control and check probe and
reference plate.
3-008
Impedance lower than 50 Ohm condition on Probe
1 with stimulate control turned on
Unable to stimulate probe 1
Impedance probe 1 too low (<50 ohms)
Turn off stimulate control, check probe placement
and leads.
3-009
Impedance lower than 50 Ohm condition on Probe
2 with stimulate control turned on
Unable to stimulate probe 2
Impedance probe 2 too low (<50 ohms)
Turn off stimulate control, check probe placement
and leads.
3-010
Impedance lower than 50 Ohm condition on Probe
3 with stimulate control turned on
Unable to stimulate probe 3
Impedance probe 3 too low (<50 ohms)
Turn off stimulate control, check probe placement
and leads.
3-011