NEUROTHERM RADIO FREQUENCY LESION GENERATOR
3-7
MODEL NT1100
®
OPERATORS MANUAL
Document 109.00
10September2012 Issue 06
Earth Leakage
Typical
Maximum Allowable
1
Enclosure leakage current
Normal
Reverse
Single Fault Condition
Normal
Reverse
40 microamps
40 microamps
40 microamps
40 microamps
100 microamps
100 microamps
500 microamps
500 microamps
2
Patient leakage current
Normal (AC)
Reverse (AC)
Single fault condition
Normal (AC)
Reverse (AC)
5 microamps
4 microamps
7 microamps
7 microamps
100 microamps
100 microamps
500 microamps
500 microamps
3
Patient Leakage current
Normal (DC)
Reverse (DC)
Single fault condition
Normal (DC)
Reverse (DC)
4 microamps
4 microamps
4 microamps
4 microamps
10 microamps
10 microamps
50 microamps
50 microamps
4
Patient Auxiliary Leakage Current
Normal (AC)
Reverse (AC)
Single Fault Condition
Normal (AC)
Reverse (AC)
4 microamps
4 microamps
6 microamps
6 microamps
100 microamps
100 microamps
500 microamps
500 microamps
5
Patient Auxiliary Leakage Current
Normal (DC)
Reverse (DC)
Single Fault Condition
Normal (DC)
Reverse (DC)
4 microamps
4 microamps
4 microamps
4 microamps
10 microamps
10 microamps
50 microamps
50 microamps
6
Patient Leakage Floating Type
Normal
Reverse
Single Fault Condition
Normal
Reverse
27 microamps
27 microamps
36 microamps
35 microamps
5000 microamps
5000 microamps
5000 microamps
5000 microamps