OPERATORS MANUAL
Document 109.00
10September2012 Issue 06
Environmental Conditions
7
Transport
Temperature
Humidity
Pressure
-10°C to 70°C
0-95% RH
140-760mmHg
Non-Condensing
(0-12,200 metres)
(0-40,000ft)
8
Storage
Temperature
Humidity
Pressure
10°C to 60°C
10 to 80% RH
520-760mmHg
(0-3000 metres)
(0-10,000ft)
9
Operating
Temperature
Humidity
10°C to 40°C
10 to 80% RH
IMPORTANT NOTICE
To comply with the Waste Electronic Equipment Regulations (WEE), at the end of the equipment’s
life, the NeuroTherm unit should be returned to the manufacturer NeuroTherm for safe disposal of
the electronic components. (The return addresses are shown in section 11-1 of this manual).
Minimising Electromagnetic Interference
Although the NeuroTherm NT1100 meets the EMC requirements for a device of this type, it is
good practice to follow certain guidelines to minimise the risk of interference between the
NeuroTherm and other devices.
1
Do not twist the cable of the NeuroTherm with those of other devices
2
Avoid putting the NeuroTherm on top of other operating equipment or other
operating equipment on top of the NeuroTherm
3
The NeuroTherm generates 480 KHz at up to 30 watts during the RF Lesion
Treatment phase. If any interference occurs to other equipment, it will be most
noticeable under this condition.
To check this, connect the 200 ohm Test Box into the machine, turn to full power in RF Lesion
Mode and observe any reading changes or interference on other equipment.
To minimise any interference, position the NeuroTherm as far away as possible from the device
being interfered with.