Electrical Safety Testing
If an Automatic or Manual Electrical Safety Analyser is used, the following settings must be used:
Machine Class:
Class II Type BF
To test the various leads of the output, use the following plugs:
Dispersive plug (4mm Safety Socket)
One of the active Plugs (Lemo 4 pin)
There is no specified EARTH REFERENCE POINT as the output is floating and could possibly
induce operational errors. If an Earth point is needed, attach onto any of the four allen cap bolts
on the rear of the machine.