7.0
Troubleshooting
Problem
The display does not come on & there
is no signal from the unit
The unit is switched on but does not
respond to commands
Battery symbol flashing: Ineffective
stimulation
Stimulation received irregularly, only
at a high intensity, or not at all
Increasing intensity causes
unpleasant sensation
Load Sense symbol appears , unit
beeping
Possible Cause
Battery discharged
Battery was incorrectly positioned
Lead not fully inserted
Broken lead
The battery is low
Faulty lead
Check your skin for lotions, pigment marks,
dry marks or other factors that could increase
resistance.
Faulty lead assembly
Electrode faulty
Poor skin/electrode contact
Solution
Replace battery
Remove battery, replace correctly
Remove plug, reinsert
Replace electrode/ lead assembly
Replace the battery
Replace lead
Slowly move electrode to locate area
where stimulus feels strongest
Moisten electrodes
Wash any oils from the skin
Check connections, replace if broken
Replace electrode
Check electrode contact with skin
14
TROUBLE
SHOO
TING
MediStim XP IM W Europe_PolyStim XP GA 290905.qxd 01/12/2010 12:49 Page 16
Содержание MediStim XP
Страница 1: ...MediStim XP IM W Europe_PolyStim XP GA 290905 qxd 01 12 2010 12 49 Page 3 ...
Страница 114: ...MediStim XP IM W Europe_PolyStim XP GA 290905 qxd 01 12 2010 12 50 Page 116 ...
Страница 115: ...MediStim XP IM W Europe_PolyStim XP GA 290905 qxd 01 12 2010 12 50 Page 117 ...
Страница 116: ...MediStim XP IM W Europe_PolyStim XP GA 290905 qxd 01 12 2010 12 50 Page 2 ...