Neuron-Spectrum (Technical Manual)
6
·
Review, store and print of the recorded traces, results of their analysis and exam
reports.
A patient stimulation can be performed with the use of stimulators built in the device.
The main condition for using the digital EEG systems is the good professional skills of
the medical staff.
1.2. Safety Measures
To ensure the safety and exclude the hazard of medical staff’s or patient’s
electric shock, the medical staff is PROHIBITED:
·
to use the digital EEG system which was mounted and installed incorrectly, without
following this manual instructions;
·
to connect digital EEG system and other high-frequency equipment to patient (it can
lead to burns in places of electrode attachment or system damage);
·
to connect another equipment (not included in delivery set) to electrode slots;
·
to open the units to make repair;
·
to work with the digital EEG system when the electronic unit, computer or other
devices used together with the stimulator are opened;
·
to connect electrodes placed on the patient to
protective ground
or other
conducting surfaces.
1.3. Specifications
Table 1. Main Specifications
Parameters
Values
EEG Channels
Number of channels
8/16/19/21*
Sampling rate
100, 200, 500, 1000, 2000,
5000 Hz
Input range
1–12000 µV
Ratio error of voltage measurement:
·
in the range from 10 up to 50 µV
·
in the range from 51 up to 12000 µV
within ±15%
within ±5%
Ratio error of time interval measurement in the range from
10 µs up to 10 s
within ±2%