Neuron-Spectrum (Technical Manual)
30
Fig. 11. Auditory stimulator electrical schematic.
5.5. Photic Stimulator Repair
Photic stimulator is examined externally for the cable failure detection. Circuit control
from the jack side is conducted according to the schematic in Fig. 12 by the device for
LED check. It is necessary to open cable jack unit and check the montage. If the
failure is not detected, open the photic stimulator unit and measure the resistance of
each cable. If the cable break is detected, see the instructions on USB cable repair
(see chapter 5.3 “Computer Interface Cable Repair (USB cable)). Detection and
elimination of LED failure can be done only by firm-manufacturer.
Fig. 12. Photic stimulator electrical schematic.