Neuron-Spectrum (Technical Manual)
26
3.2. Setting-up Procedures
Operating Limitations:
·
Ambient temperature is from +10 to +35
°
С
.
·
Relative humidity is from 30% to 85%.
·
Atmospheric pressure is from 70 to 106 kPa.
Before power supply switching make sure that digital EEG electronic unit and
computer equipment cases have no apparent mechanical failures which can represent
a danger.
Power Supply Switch on and Digital EEG System Test
Digital EEG system power supply switch on is done by pressing the “Power” key of
your PC. The electronic unit has no power supply switch and is constantly connected
to PC. Power supply switch on occurs after PC operational system loading and
Neuron-Spectrum
program start.
Troubleshooting
The list of some possible troubles and the ways of their removal is given in Table 6.
Table 6. Possible Troubleshooting
Trouble symptom
Cause
Way of Removal
When PC switching on, the
indicators on the PC front panel do
not glow.
Incorrect PC connection to the
mains.
Check the PC connection accuracy
to the mains.
PC operational system load is not
carried out properly.
Faultiness of PC hardware or
software.
Make PC reload. In case the
trouble was not removed, address
to PC firm-supplier.
Neuron-Spectrum.NET
software
doesn’t start.
Digital EEG system software is not
installed or is installed incorrectly.
Check whether the digital EEG
system software is set up. Reinstall
digital EEG system software using
distribution disc.
When switching on the monitoring
mode, the EEG signal can not be
registered.
Digital EEG system is connected
incorrectly.
1. Check whether digital EEG
system is connected.
2.e Consult the manufacturer.
When starting the program, the
message “Database connection
error” can appear on the screen.
1. The directory with patient exam
files was renamed or replaced.
2. The directory with patient exam
files is inaccessible.
1. Check the connection to local
network. Make sure that the
computer with patient files directory
is connected to local network and
switched on.
2. Restore the name and the
directory placement.