Treatment Timer
To change the treatment time, press the program button once again.
The NeuroMove® can time therapy sessions. The selectable treatment times are
from 10 - 90 minutes, in ten-minute increments. The default time is
continuous. The unit will automatically shut off after the selected amount of
time. In continuous mode, the unit never turns off.
Treatment Time:
Continuous
To increase the treatment time, press the Up button. To decrease, press the
Down button.
Set Defaults
To set the factory defaults, press the program button once again.
If any of the parameters are mistakenly set, the NeuroMove® provides a way to
return to factory default conditions.
Set Defaults?
↓
To set defaults press the Down button. The display will show “DefaultsSet”.
Press the programming button again to exit the Change Parameters
subfunction.