Operational Modes
To change the operational mode, press the program button once again.
The NeuroMove® has four different operational modes.
Stroke Rehabilitation
This is the normal mode for the NeuroMove®. It has good low-level
sensitivity and a wide dynamic range. The range is 1.25-1000 micro-
volts (58 decibels). It is intended primarily for stroke rehabilitation.
SCI (Spinal Cord Injury)
This mode is similar to Stroke Rehabilitation but has more low-level
sensitivity and less range. It is intended primarily for spinal cord
injury
rehabilitation. There are two sensitivity ranges available: 0.25-25
microvolts (40 decibels) and 0.50-50 microvolts (40 decibels). See
Change Parameters programming below for range selection
procedure.
Manual
This mode is similar to Stroke Rehabilitation, but requires manual
setting of the threshold.
Stim Only
In this mode, the unit acts like a stimulator only. See “Operational
Modes Descriptions” below.
Mode:
Stroke Rehab
Use Up and Down buttons to select the desired mode.