8. USER INTERFACE & DISPLAY
8
6
7
®000
3
5
4
ltem Button Name
Function
Single press, motor will start as per selected rpm and
1
Start
time if lid closed. Startbutton works only if lid closed.
Motor start or stop, indicates with LED
Single press stop/open button, will stop motor as per
2
Stop/op en
defined DCC time if motor is running. Second press stop/
open button, will open the lid if motor stopped. Also use
as lid open function if centrifuge motor is not running.
Single press, RPM will be display like 4500 and RPM LED
3
RPM/RCF
will blink. Lang Press, RPM/RCF will toggle display to
RPM to RCF.
RPM , RCF mode also indicate via LED on panel.
4
Time
Single press, Time will be display like 99.00
Timer can be set from1 to 99 min.
Short press of {+}Plus and {-}Minus buttons will increase
5
lncrement/
and decrease speed, time, ACC, DCC values normally.
Decrements
Lang press of these keys would change the parameters
quicker
First Single press, Acceleration will display on time
Acceleration
segment like ACC1. We can change acceleration using
6
{+} and {-} buttons. (1 to 9).
Second Single press, Deceleration will display on time
Deceleration
segment like DCC1. We can change Deceleration using
{+} and {-} buttons. (1 to 9).
7
Rotor
To select among multiple rotors listed below
Single press Pedit, SPEED/ TIME/ ACC/ DCC will blink
one after one 5 times on respective display and user can
8
Pedit
change blinking parameters by using {+}Plus and {-}Minus
key. In PEdit mode user can also press SPEED/TIME/ACC/
DCC and change its value.