![Neuation iFUGE L400P Скачать руководство пользователя страница 11](http://html1.mh-extra.com/html/neuation/ifuge-l400p/ifuge-l400p_product-user-manual_1675691011.webp)
L30-134
Rotor 9
6x50 ml
4500
2644
L30-137
Rotor 10
24x 15 ml
4500
2852
L30-131
Rotor 11
16 x 15 ml
4500
3144
L30-132
Rotor 12
12 x 15 ml
4500
3144
L30-133
Rotor 13
8x 15 ml
4500
2644
L30-138
Rotor46
30x 15 ml
4500
2729
START&STOP
The centrifugation can be started by pressing start button once the desired
operation set. The operation can be stopped by pressing stop button once. To
open the lid press stop/open key once after the rotor comes to stop.
10. MAINTENANCE AND CLEANING
•
The rotor and the outside of the centrifuge should be cleaned regularly with a
mild wet (with water) cloth.
•
Ensure that while cleaning the unit is not plugging in.
•
Wear protective glove & safety glass while operating & cleaning the device.
•
The brushless motor in the centrifuge requires no routine maintenance. Any
required service should be performed by authorized, qualified personnel
only. Repairs performed by unauthorized personnel may void the warranty.
•
Always keep the centrifuge housing, rotor chamber and rotor clean. All parts
should be wiped down periodically with a soft cloth. For more thorough
cleaning, use a neutral cleaning agent (Ph between 6 and 8) and clean with a
soft cloth. Exclusive amounts of liquid should be avoided.
Note:
Liquid shou/d not come into contact with the motor.
•
After cleaning, ensure that all parts are dry before re-use.
•
Regularly cleaning of the rotor is important.
•
lf the rotor chamber needs cleaning, clean with cloth or sponge moistened
with a neutral detergent solution.
•
Do not place the rotor into the cleaning solution.
•
lf corrosive, toxic or pathogenic bacteria are accidentally spilled in the rotor
or rotor chamber the centrifuge must be decontaminated throughly.
11. TROUBLESHOOTING
This centrifuge has a seif - diagnostic function. lf a problem occurs, an
error/warning code will be displayed on the display screen and the operator can
determine the malfunction with the warning code below.
II