Neuation iFUGE L400P Скачать руководство пользователя страница 10

PRG: 

Press PRG button to select the program from 1-99. The different parameters can 

be set in each programm for quick selection. To set parameters in each program 

follow the parameters (Speed, time & Acc/Dcc) as described above. To select from 

programs press using +/- button for increment or decrement. All the parameters in 

the program saves automatically once the blinking stops. The saved programs also 

can be edited by following the instruction of setting Speed, time & Acc/Dcc as 

described above. 

Quick Spin: 

Press hold the quick spin button for immediate centrifugation of the samples up to 

previous run RPM. Releasing the hold will stop of the rotor gradually. 

Pedit: 

Press  PEdit  button  to  enter  all  the  different  parameters  - Speed,  Timer, 

acceleration  & deceleration automatically one after the other. These parameter 

will be  asked by  the instrument sequentially  to be entered if  required  to be 

changed  (if not edited the display  value will set).  These parameters will save 

automatically. 

Rotor: 

Press rotor button to select rotors as listed below. The rotor numbers can be 

changed by using+/- for increment or decrement of the rotor number. 

Swing Out Rotors 

Rotor 

Rotor No. 

Max. Volume 

Max. RPM 

Max. RCF 

L30-130 

Rotor 1 

4x 100 ml 

4500 

3164 

L30-129 

Rotor 2 

4x 50 ml 

4500 

3485 

L30-128 

Rotor 3 

6x 50 ml 

4500 

3530 

L30-125 

Rotor4 

16 x 15 ml 

4500 

3485 

L30-126 

Rotor 5 

12 x 15 ml 

4500 

3485 

L30-140 

Rotor 25 

32 x 10 ml 

4500 

3164 

L30-124 

Rotor 26 

2 x Microplates 

4500 

2377 

Fixed Angle Rotors 

Rotor 

Rotor No. 

Max. Volume 

Max. RPM 

Max. RCF 

L30-136 

Rotor 7 

4x 100 ml 

4500 

2559 

L30-135 

Rotor 8 

4x 50 ml 

4500 

2644 

II 

Содержание iFUGE L400P

Страница 1: ...NEUATION TECHNOLOGIES iFUGE L400P PRODUCT USER MANUAL...

Страница 2: ......

Страница 3: ...FETY PRECAUTIONS 2 7 INSTALLATION 3 Location and Mounting 8 USER INTERFACE AND DISPLAY 4 9 OPERATING THE CENTRIFUGE 5 Switch on the Centrifuge Speed Setting Time Setting Acc Dcc PRG Quick Spin PEdit p...

Страница 4: ......

Страница 5: ...this user manua carefully This user manua is intended to assist with the operation and care of the unit only and not its repair For repairs p easecontactthesupplier 3 SALIENTFEATURES Centrifuge hasfol...

Страница 6: ...ely fastened lf the centrifuge makes unusual noise during operation the rotor fitment needs to be checked Switch OFFthe deviceimmediately by pressing STOP check fitment fastenitweil Never movethecentr...

Страница 7: ...afety wehave provided protectiveearthing with power supply Make sure powersupply is earthened 7 INSTALLATION The Laboratory centrifuge is supplied in packaging box Open the box remove the packaging an...

Страница 8: ...e Single press Time willbe display like 99 00 Timer canbe set from1 to 99 min Short press of Plus and Minusbuttons will increase 5 lncrement and decrease speed time ACC DCC values normally Decrements...

Страница 9: ...g button for increment button for decrement Thevalueswillautomatically savesafterscreenstopsblinking TimerSetting Press the timer button to set timer Time can be set by for increment for decrement The...

Страница 10: ...fferent parameters Speed Timer acceleration deceleration automatically one after the other These parameter will be asked by the instrument sequentially to be entered if required to be changed if not e...

Страница 11: ...d personnel only Repairsperformedby unauthorizedpersonnelmay voidthewarranty Always keep the centrifuge housing rotor chamber and rotor clean All parts should be wiped down periodically with a soft cl...

Страница 12: ...the rotor or rotor motor contact service exceptional motor is cause for run representative running noise noise Rotor damaged Remove change rotor Display error Loose connection of display Contact servi...

Страница 13: ...he product or use of products as weil as any supervision required for safety lf requested the products must be returned to the distributor in weil packed and insured manner and all shippingchargesmust...

Страница 14: ...quipment WEEE According to these regulations any devices suppliedafter 13 06 05 inthe business to businesssphere to whichthis product is assigned may no langer be disposed off in municipal or domestic...

Страница 15: ......

Страница 16: ...Neuation Technologies Pvt Ltd Plot No 15 GIDC Electronic Park SEZ Kolavada Road Gandhinagar 382026 Gujarat India Website www neuation com...

Отзывы: