ERROR
PROBLEM
SOLUTION
No main power
Power check
&
proper plug-in of
connection.
mains cable at both ends.
No display
Power failure
Check the mains fuse of the lab.
lmproper connection.
Connect adaptor properly.
a
Lid not closed correctly.
Close lid correctly.
Error with lid closing and
Contact service.
opening mechanism.
Rotor not loaded
Load rotor symmetrically
&
restart
Err 55
symmetrically.
centrifuge.
Centrifuge lid
Rotor is still spinning.
Wait for the rotor to come to a stop.
cannot be opened
Power failure
Emergency lid release after
rotor stops
Rotor not loaded
Load rotor symmetrically
&
Centrifuge shakes
symmetrically.
restart operation
during
Either a broken tube,
Replace broken tube. For damaged
acceleration
&
damage to the rotor or
rotor/motor contact service
exceptional
motor is cause for run
representative.
running noise
noise.
Rotor damaged.
Remove
&
change rotor
Display error
Loose connection of display.
Contact service representative.
Latch damaged, Latch
Contact service
Err 1
jammed
representative.
Motor stuck or incorrect Turn OFF the centrifuge, Check rotor
Err 52
operating voltage
fitment or apply correct 230VAC ±
l0VAC operating voltage
Power tripping
Cable not fit properly.
Remove cable and connect properly.
Last run memory
Turing ON centrifuge
Maintain 3 seconds gap Between
not displayed
immediately after
switch OFF and switching
ON
again.
turning it OFF.
Switch off centrifuge and then
System gets hang
Electronics error.
switch it
ON
again. lf the error still
shows, contact service representative.
IMPORTANTNOTE:
•
lf system get hangs or gets heated due to over current, switch OFF & switch
ON (restart) the centrifuge and check it again.