2.10.4 Printer Format Selection
Different print formats may be selected. Press the ‘PrSF’ F3
key to select a different format. Press the ‘
←
’ and ‘
→
’ keys to move
through the selections. The selections are listed :
TAB
Fix horizontal space
LF
Line feed
FF
Form feed
CR
Carriage
return
DW
Double width print
DCLR
Reset double width print
HW
Condensed
HCLR
Reset condensed print
EMP
Emphasize print
ECLR
Reset emphasize print
DST
Double strike
CLR
Reset double strike
MarR
Right margin set
BS
Back
space
RST
Reset to power on mode
MarL
Left margin set
PRT
Print 16 characters
2.10.5 Pacemaker Battery Current Test
The EXPMT 2000 will measure the battery current of the
pacemaker to be tested. The Netech 652-Battery Adapter is required to
perform this measurement.
A.
Connect the battery adapter to the pacemaker and a 9 Volt
battery.
B.
Connect the banana jacks of the battery adapter to the
VENT connectors of the A.V. INPUT section on the side of
the EXPMT 2000.
C.
Press the Bat ‘F4’ key to perform the measurement.
EXPMT 2000 User Manual
27