There are three levels of soft key menus in the Main Pacer Test
Menu.
Level 1: PPM Wdth Amp Ref
Level 2: Sens Imm AV-D Enr
Level 3: Type Aut Pri Aux
Press either the ‘
←
’ or ‘
→
’ key to move between the three soft key
menu Levels.
2.5.1 Selecting Pacemaker Manufacturer
A. Move the PACER INPUT SELECTOR switch on the side
to the TRANS. INPUT position.
B. Turn the EXPMT 2000 instrument on.
C. Press the ZOLL ‘F1’, HP ‘F2’, MDE ‘F3’, or LIFE ‘F4’ key.
D. Press either the ‘
←
’ or ‘
→
’ key to move to the next menu
level and then press the ‘LF-12’ key to select LifePak 12
models or the ‘Others’ key to select all other
manufacturers.
E. A test function is selected using ‘F1’, ‘F2’, ‘F3’, or ‘F4’
keys.
F. Press either the ‘
←
’ or ‘
→
’ key to reach Level 2 and
Level 3 menus.
2.5.2
Selecting 50 Ohm Fixed Test Load
A.
Connect the transcutaneous pacer to the TRANS INPUT
fixed 50 OHMS input connectors (see the next section for
variable loads).
B.
Turn the EXPMT 2000 on.
C.
Turn the pacer on.
D.
Select the desired pacer manufacturer.
E.
Select the desired parameter to be measured (PPM,
Wdth, Amp, Ref) using ‘F1’, ‘F2’, ‘F3’, or ‘F4’ keys.
F.
The EXPMT 2000 will now display the measured
parameters.
G.
Press either the ‘
←
’ or ‘
→
’ key to reach Level 2 and
Level 3 menus.
EXPMT 2000 User Manual