Controls & Indicators 

Item 

Name 

Description 

Display 

A 3½ Digit LCD indicates the result of the measurement being made 

 

 

LEDs 

 

3 red LEDS indicate the selected test mode. 

Mode Switch 

Three-position slide switch to select the test measurement. A lit red 

LED indicates the selection. 

Ground 

Switch 

A switch to momentarily open the ground connection to the test 

receptacle 

Polarity 

Switch 

A three-position switch to select the normal and reverse polarity of 

the test receptacle. The center position shuts the power off to the test 

receptacle. 

Neutral 

Switch 

A switch to momentarily open the neutral line to the test receptacle. 

Load Selector 

Switch 

A slide switch on the right side of the instrument to select either the 

AAMI or the IEC 601-1 test load. 

Test 

Receptacle 

A 20 AMP power receptacle for the Device Under Test 

Power Cord 

The power cord supplies power to the analyzer and to the Device 

Under Test 

10 

Case Ground 

A red jack is provided for connecting the test lead to the case 

(chassis) of the Device Under Test. The Cord Resistance and 

Case Leakage are measured using this cable 

11 

Test Jacks 

A calibrated output is provided at the test jack for resistance and 

current functions. This performs a self-check of the LKG 601 

12 

Zero 

Adjustment 

This recessed adjustment allows for the zeroing of the display to 

eliminate test cable resistance variations 

13 

Fuse 

15 Ampere fuse is in an accessible fuse holder for 110 Volt 

model.  The 230 Volt model uses a 10-Ampere fuse 

Содержание 510-110

Страница 1: ...LKG 601 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www NetechCorp us http www PressureMeter com LKG 601 ELECTRICAL SAFETY ANALYZER INSTRUCTION MANUAL...

Страница 2: ...pair and Recalibration 5 Models and Part Numbers 6 General Overview Introduction 8 Key Features 8 Specifications 9 Instrument Familiarity 10 Controls Indicators 11 Preparation for use 12 Operating Ins...

Страница 3: ...otice and without incurring any obligation Netech guarantees availability of service parts for 5 years after the manufacture of the unit is discontinued The warranty is void if you elect to have the u...

Страница 4: ...st procedures are found to meet those specifications when it was shipped from the factory Calibration LKG 601 is calibrated using standards traceable to National Institute of Standards and Technology...

Страница 5: ...return or refund Only products returned within 60 days from the date of original purchase are eligible for refund credit In order to receive a partial refund credit of a product purchase price on a s...

Страница 6: ...aging or handling Use the original carton and packaging material for shipment Returns for Refund Credit A Return Material Authorization RMA number must be obtained from our service or customer service...

Страница 7: ...Hard case 510 220 Electrical Safety Analyzer 220 V with Hard case 510 110 220 Electrical Safety Analyzer 110 V and 220V Operation with Hard Case Standard Accessories Item Description Picture 503 Test...

Страница 8: ...ects the test to be performed The user may select either the AAMI ESI 1993 or the IEC 601 1 test load to compensate for high frequency components in the measurement Accurate resistance measurements ar...

Страница 9: ...TO 1999 Milli Ohms Accuracy 1 of reading 1 LSD DEVICE CURRENT 0 to 19 99 Amps Accuracy 2 Full Scale 1 LSD TEST RECEPTACLE Hospital Grade 110 VAC 15 Amp or 220 VAC 10 Amp The rocker polarity switch se...

Страница 10: ...User Manual LKG 601 10 Instrument Familiarity 1 3 6 5 2 4 7 9 8 13 11 10 12...

Страница 11: ...pen the neutral line to the test receptacle 7 Load Selector Switch A slide switch on the right side of the instrument to select either the AAMI or the IEC 601 1 test load 8 Test Receptacle A 20 AMP po...

Страница 12: ...n of the dual voltage LKG 601 Before proceeding with any measurement become familiar with the measurements and the function selector switch During the initial setup the last digit of the LCD will chan...

Страница 13: ...to CORD RESISTANCE Plug the Device Under Test into the test outlet Using the test lead provided connect the chassis of the Device Under Test to the Case Ground jack on the LKG 601 The display will in...

Страница 14: ...e case of the DUT Device under test The leakage measurement is made when the Ground switch is momentarily opened The LCD will display the measurement in Micro Amps Both Earth and Case Leakage Current...

Страница 15: ...indicating that the cord circuit is open 2 Move the Mode selector switch to all three positions and test the LED indication 3 Check the continuity between receptacle ground and case ground using the...

Страница 16: ...switch selects leakage current measurement the AAMI LOAD is connected in front of the INPUT AMPLIFIER and the voltage developed across it from the leakage current is measured Current Source and Amplif...

Страница 17: ...libration is recommended If the LKG 601 is returned to Netech for recalibration before the first year warranty expires Netech will provide a second year warranty The warranty is void if the LKG 601 is...

Страница 18: ...A number follow the link http www NetechCorp us and fill in the required information Email service netechcorp us or call 800 547 6557 US Canada International 631 531 0100 When shipping units to the fa...

Страница 19: ...dapter for 220 Volt devices CAUTION LKG 601 dual voltage model PN 510 110 220 designed for use in 110 Volt Ac powered equipments or 220 AC powered equipments For 220 volt devices make sure the device...

Страница 20: ...User Manual LKG 601 20 Appendix This section is intentionally left blank for future use...

Отзывы: