LKG 601
Netech Corporation
110 Toledo St, Farmingdale, NY 11735,
Controls & Indicators
Item
Name
Description
1
Display
A 3½ Digit LCD indicates the result of the measurement being made
2
LEDs
3 red LEDS indicate the selected test mode.
3
Mode Switch
Three-position slide switch to select the test measurement. A lit red
LED indicates the selection.
4
Ground
Switch
A switch to momentarily open the ground connection to the test
receptacle
5
Polarity
Switch
A three-position switch to select the normal and reverse polarity of
the test receptacle. The center position shuts the power off to the test
receptacle.
6
Neutral
Switch
A switch to momentarily open the neutral line to the test receptacle.
7
Load Selector
Switch
A slide switch on the right side of the instrument to select either the
AAMI or the IEC 601-1 test load.
8
Test
Receptacle
A 20 AMP power receptacle for the Device Under Test
9
Power Cord
The power cord supplies power to the analyzer and to the Device
Under Test
10
Case Ground
A red jack is provided for connecting the test lead to the case
(chassis) of the Device Under Test. The Cord Resistance and
Case Leakage are measured using this cable
11
Test Jacks
A calibrated output is provided at the test jack for resistance and
current functions. This performs a self-check of the LKG 601
12
Zero
Adjustment
This recessed adjustment allows for the zeroing of the display to
eliminate test cable resistance variations
13
Fuse
15 Ampere fuse is in an accessible fuse holder for 110 Volt
model. The 230 Volt model uses a 10-Ampere fuse