User Manual LKG 601
12
Preparation for Use
The LKG 601 includes a test lead (PN: 503) for measuring Cord Resistance and Case
Leakage. No other cables are required to perform measurements.
There are three models of LKG 601 available. Prior to using the device identify the Part
number and the model purchased.
The following are the part numbers and the operational voltages.
1. PN: 510-110, for 110 Volt devices only.
2. PN: 510-220, for 220 Volt devices only.
3. 510-110-220 dual voltage operation.
Please refer to the ADDENDUM for the preparation, setup and safe operation of the
dual voltage LKG 601.
Before proceeding with any measurement become familiar with the measurements and
the function selector switch.
During the initial setup the last digit of the LCD will change gradually due to the time
constant of the RMS to DC converter. The display will stabilize to zero in a few
seconds.
CAUTION
Make sure that the power requirements of the Device Under Test are within the
power ratings of the LKG 601, 15 amps at 110 volts and 10 amps at 230 volts.
Do not leave the Device Under Test continuously turned on and connected to the
LKG601
.
Before testing 220 Volt devices make sure the LKG 601 is plugged to a 220 Volt
out let using the adapter and connect the 220 Volt device to the LKG 601 using
the universal adapter supplied