User Manual                                    LKG 601 

 

 

12 

 

Preparation for Use 

 

The LKG 601 includes a test lead (PN: 503) for measuring Cord Resistance and Case 

Leakage. No other cables are required to perform measurements.  

 

There are three models of LKG 601 available. Prior to using the device identify the Part 

number and the model purchased.  

 

The following are the part numbers and the operational voltages.  

 

1.  PN: 510-110, for 110 Volt devices only. 

2.  PN: 510-220, for 220 Volt devices only. 

3.  510-110-220 dual voltage operation.  

 

 

Please refer to the ADDENDUM for the preparation, setup and safe operation of the 

dual voltage LKG 601.  

 

Before proceeding with any measurement become familiar with the measurements and 

the function selector switch. 

 

During the initial setup the last digit of the LCD will change gradually due to the time 

constant of the RMS to DC converter.  The display will stabilize to zero in a few 

seconds. 

 

CAUTION 

 

Make sure that the power requirements of the Device Under Test are within the 

power ratings of the LKG 601, 15 amps at 110 volts and 10 amps at 230 volts. 

 

Do not leave the Device Under Test continuously turned on and connected to the 

LKG601

 

Before testing 220 Volt devices make sure the LKG 601 is plugged to a 220 Volt 

out let using the adapter and connect the 220 Volt device to the LKG 601 using 

the universal adapter supplied

 

 

Содержание 510-110

Страница 1: ...LKG 601 Netech Corporation 110 Toledo St Farmingdale NY 11735 http www NetechCorp us http www PressureMeter com LKG 601 ELECTRICAL SAFETY ANALYZER INSTRUCTION MANUAL...

Страница 2: ...pair and Recalibration 5 Models and Part Numbers 6 General Overview Introduction 8 Key Features 8 Specifications 9 Instrument Familiarity 10 Controls Indicators 11 Preparation for use 12 Operating Ins...

Страница 3: ...otice and without incurring any obligation Netech guarantees availability of service parts for 5 years after the manufacture of the unit is discontinued The warranty is void if you elect to have the u...

Страница 4: ...st procedures are found to meet those specifications when it was shipped from the factory Calibration LKG 601 is calibrated using standards traceable to National Institute of Standards and Technology...

Страница 5: ...return or refund Only products returned within 60 days from the date of original purchase are eligible for refund credit In order to receive a partial refund credit of a product purchase price on a s...

Страница 6: ...aging or handling Use the original carton and packaging material for shipment Returns for Refund Credit A Return Material Authorization RMA number must be obtained from our service or customer service...

Страница 7: ...Hard case 510 220 Electrical Safety Analyzer 220 V with Hard case 510 110 220 Electrical Safety Analyzer 110 V and 220V Operation with Hard Case Standard Accessories Item Description Picture 503 Test...

Страница 8: ...ects the test to be performed The user may select either the AAMI ESI 1993 or the IEC 601 1 test load to compensate for high frequency components in the measurement Accurate resistance measurements ar...

Страница 9: ...TO 1999 Milli Ohms Accuracy 1 of reading 1 LSD DEVICE CURRENT 0 to 19 99 Amps Accuracy 2 Full Scale 1 LSD TEST RECEPTACLE Hospital Grade 110 VAC 15 Amp or 220 VAC 10 Amp The rocker polarity switch se...

Страница 10: ...User Manual LKG 601 10 Instrument Familiarity 1 3 6 5 2 4 7 9 8 13 11 10 12...

Страница 11: ...pen the neutral line to the test receptacle 7 Load Selector Switch A slide switch on the right side of the instrument to select either the AAMI or the IEC 601 1 test load 8 Test Receptacle A 20 AMP po...

Страница 12: ...n of the dual voltage LKG 601 Before proceeding with any measurement become familiar with the measurements and the function selector switch During the initial setup the last digit of the LCD will chan...

Страница 13: ...to CORD RESISTANCE Plug the Device Under Test into the test outlet Using the test lead provided connect the chassis of the Device Under Test to the Case Ground jack on the LKG 601 The display will in...

Страница 14: ...e case of the DUT Device under test The leakage measurement is made when the Ground switch is momentarily opened The LCD will display the measurement in Micro Amps Both Earth and Case Leakage Current...

Страница 15: ...indicating that the cord circuit is open 2 Move the Mode selector switch to all three positions and test the LED indication 3 Check the continuity between receptacle ground and case ground using the...

Страница 16: ...switch selects leakage current measurement the AAMI LOAD is connected in front of the INPUT AMPLIFIER and the voltage developed across it from the leakage current is measured Current Source and Amplif...

Страница 17: ...libration is recommended If the LKG 601 is returned to Netech for recalibration before the first year warranty expires Netech will provide a second year warranty The warranty is void if the LKG 601 is...

Страница 18: ...A number follow the link http www NetechCorp us and fill in the required information Email service netechcorp us or call 800 547 6557 US Canada International 631 531 0100 When shipping units to the fa...

Страница 19: ...dapter for 220 Volt devices CAUTION LKG 601 dual voltage model PN 510 110 220 designed for use in 110 Volt Ac powered equipments or 220 AC powered equipments For 220 volt devices make sure the device...

Страница 20: ...User Manual LKG 601 20 Appendix This section is intentionally left blank for future use...

Отзывы: