- 32 -
Error Message
Type
Indication
Corrective Action
COOL (Steady)
Status
COOL (Flashing) Status
PUMP
Status
LOCAL
Status
REMOTE
Status
FLUID LEAK
Failure
LEVEL FAULT
Failure
(Steady)
LEVEL FAULT
Failure
(Flashing)
Controller is asking for the cooling
solenoid to turn on.
Controller is maintaining tempera-
ture control within the proportional
band by asking the solenoid valve
to turn on and off.
The pump is on either locally or
remotely.
The unit is on and in the local
mode.
The unit is on and in the remote
mode.
The unit drip pan is accumulating
facility water or recirculating fluid.
Heater is disabled, unit will
continue to operate. If unit is off
the unit will not turn on.
Unit is off and will not turn on.
Fluid level went from an ADD
FLUID warning to a LEVEL
FAULT within three minutes.
None.
None.
None.
None.
None.
Immediately locate the cause of the leak and
turn off the unit if a failure is eminent. Repair
leak or replace unit.
Add Fluid until the ADD FLUID led goes out
when reset.
Check for fluid leak in system. The fault will
reset once the reservoir is filled past the
ADD FLUID level.