- 13 -
Operators Loop
When the controller is first powered it goes through a short self test and then
enters the Operator's Loop, displaying the reservoir temperature as measured
by either the internal or external RTD probe.
By pressing the NEXT key the controller will step through the menu shown
below.
YES
key increments setpoint.
NO
key decrements
setpoint. See Changing a Value on page 20.
The display will flash while the setpoint is being
changed. The new value is not used by the
controller until the
ENTER
key is depressed.
Displays the flow rate of the individual circulation
loops in liters/minute. If the flow alarm is set to
zero (see Setup Loop)
FLOW OFF
will be
displayed instead.
Figure 2 Operator's Loop
This displays the resistivity of the bath fluid in
megohm-cm. Tenths will be display up to 9.9 meg.
10 - 20 meg will not display tenths. Resistivity is
maintained at or above the setpoint.