background image

 

 

1

YOUR NEW NEOPIPETTA 

The neoPipetta has been carefully designed and manufactured as a lightweight, cordless Pipette Filler. neoPipetta 

offers simple, efficient pipetting performance and maximum pipetting comfort. 

It can be used with all plastic or glass pipettes between 0.1-100ml. 

2

Charging the neoPipetta

 

Before the first use the battery must be charged for 10 hours. Allow the unit to charge anywhere from 2 to 3 hours 

for all subsequent recharging. More than 8 hours of intermittent use is typical when fully charged. 

 

Before connecting the AC-adapter to the electrical outlet, check that its voltage setting corresponding to 

the voltage of your electrical outlet. 

 

Connect the AC-adapter to the electrical outlet and to the electrical fitting at the base of the handle of 

the neoPipetta   

Note:

 The neoPipetta should only be charged with the AC-adapter provided. 

Arrows on the top indicate the power situation. When need to be charged, it shows red. 

Special circuitry prevents overcharging of the battery and allows the neoPipetta to be charged or partially charged 

at any convenient time. 

2.1. Electrical specifications 

Battery:          3.6 V, 1100mA, rechargeable lithium-ion battery   

                        Charging time 10 hours for completely drained batteries 

AC-adapter:    Input AC 100-240V 

                        Output 4.2v 500mA 

                         

 

3.    NEOPIPETTA OVERIVIEW (Fig. 1) 

 

 

Fig.1 

Nose cone 

Nose cone 

Case 

Aspirating button 

Dispensing button 

The battery compartment lid 

Speed adjustment button 

Recharge socket 

Silicone adaptor 

Power display 

Содержание neoPipetta D-6019

Страница 1: ...neoPipetta D 6019 User Manual...

Страница 2: ...ne and silicone adaptor 2 4 4 Filters 2 5 OPERATING THE NEOPIPETTA 2 5 1 Setting the operating speed 2 5 2 Inserting a serological pipet 2 5 3 Aspiration 3 5 4 Dispensing 3 5 5 Other pipetting recomme...

Страница 3: ......

Страница 4: ...ltage of your electrical outlet Connect the AC adapter to the electrical outlet and to the electrical fitting at the base of the handle of the neoPipetta Note The neoPipetta should only be charged wit...

Страница 5: ...ptor exposes a hydrophobic filter 0 45 micron located at the top of the nose piece The filter protects the neoPipetta from accidental over aspiration It is strongly recommended to use only the origina...

Страница 6: ...ument in an atmosphere with a danger of explosion Never use force with the neoPipetta Should the instrument fail to function properly stop pipetting and check the Trouble Shooting section 6 MAINTENANC...

Страница 7: ...the battery compartment exposing a 2pin connector on the PCB Connect the 2 pin plug of the new Lithium ion battery to the exposed 2 pin connector on the PCB The side of the plug with the exposed meta...

Страница 8: ...d in the wrong orientation Failure of the Printed Circuit Board Replace battery Refer to section 6 3 of manual to plug battery in correctly Call the manufacturer Tech Support 9 WARRANTY INFORMATION Th...

Страница 9: ...neoLab Migge GmbH Rischerstr 7 9 69123 Heidelberg Germany Tel 49 06221 8442 0 Fax 49 06221 8442 9933 Email info neolab de www neolab de...

Отзывы: