5.3. Aspiration 

 

Select a speed and fill the pipet by pressing the aspirating button. Fill the pipet just above the desired 

level, check level by the meniscus. 

 

Remove the serological pipet from the sample. 

 

Adjust the volume of the sample by pressing the dispensing button until the meniscus is at the required 

level. 

5.4. Dispensing 

The neoPipetta offers two different dispensing modes: 

 

If  the  dispensing  button  is  pressed  only  partially  to  a  point  where  a  check  can  be  felt,  the  dispensing 

function will be based on the force of gravity and the liquid will simply flow out. 

 

Pressing the dispensing button further in will engage the motor and activates the blow-out function. 

Note:

 Finger pressure can be varied to fine-tune the speed within each mode. 

5.5 Other pipetting recommendations 

 

Use the instrument within the limits of its chemical resistance. 

 

Never use with highly flammable liquids. 

 

Observe the reagent manufacturer’s safety information. 

 

Never use the instrument in an atmosphere with a danger of explosion. 

 

Never use force with the neoPipetta. 

 

Should the instrument fail to function properly, stop pipetting and check the Trouble Shooting section. 

 

 

 

6. MAINTENANCE 

The neoPipetta has been designed for easy in-house maintenance. 

6.1. Decontamination 

To decontaminate, lightly spray the outside of the neoPipetta with the decontamination spray and wipe dry with a 

clean cloth.   

6.2. Changing the nose cone, silicone adapter and filter (Fig. 2) 

To change the nose cone, silicone adapter and filter firmly hold the nose cone by the ribbed collar, rotate the nose 

cone counter-clockwise until completely loosen and pull out the nose cone and the silicone adapter. The filter can 

be removed by simply pulling it out once the nose cone and the silicone adapter have been disengaged. Position 

the new filter with the colored side upwards. To reassemble, insert the adapter and nose cone and twist the nose 

cone gently clockwise. 

Note:

 It is recommended to check and clean the silicone adapter every six months, or when necessary, to prevent a 

loss  of  vacuum.  If  there  is  a  loss  of  vacuum,  please  contact  the  manufacturer  or  your  local  distributor.

Содержание neoPipetta D-6019

Страница 1: ...neoPipetta D 6019 User Manual...

Страница 2: ...ne and silicone adaptor 2 4 4 Filters 2 5 OPERATING THE NEOPIPETTA 2 5 1 Setting the operating speed 2 5 2 Inserting a serological pipet 2 5 3 Aspiration 3 5 4 Dispensing 3 5 5 Other pipetting recomme...

Страница 3: ......

Страница 4: ...ltage of your electrical outlet Connect the AC adapter to the electrical outlet and to the electrical fitting at the base of the handle of the neoPipetta Note The neoPipetta should only be charged wit...

Страница 5: ...ptor exposes a hydrophobic filter 0 45 micron located at the top of the nose piece The filter protects the neoPipetta from accidental over aspiration It is strongly recommended to use only the origina...

Страница 6: ...ument in an atmosphere with a danger of explosion Never use force with the neoPipetta Should the instrument fail to function properly stop pipetting and check the Trouble Shooting section 6 MAINTENANC...

Страница 7: ...the battery compartment exposing a 2pin connector on the PCB Connect the 2 pin plug of the new Lithium ion battery to the exposed 2 pin connector on the PCB The side of the plug with the exposed meta...

Страница 8: ...d in the wrong orientation Failure of the Printed Circuit Board Replace battery Refer to section 6 3 of manual to plug battery in correctly Call the manufacturer Tech Support 9 WARRANTY INFORMATION Th...

Страница 9: ...neoLab Migge GmbH Rischerstr 7 9 69123 Heidelberg Germany Tel 49 06221 8442 0 Fax 49 06221 8442 9933 Email info neolab de www neolab de...

Отзывы: